Optimizing 3D EPI for rapid T1 -weighted imaging
Abstract
Purpose: To investigate the use of 3D EPI for rapid T1 -weighted brain imaging, focusing on the RF pulse's influence on the contrast between gray and white matter.
Methods: An interleaved 3D EPI sequence use partial Fourier and CAIPIRINHA sampling was used to acquire T1 -weighted brain volumes with isotropic resolution, low echo times, and low geometric distortions. Five different RF pulses were evaluated in terms of fat suppression performance and gray-white matter contrast. Two binomial RF pulses were compared to a single rectangular (WE-rect) RF pulse exciting only water, and two new RF pulses developed in this work, where one was an extension of the WE-rect, and the other was an SLR pulse. The technique was demonstrated in three clinical cases, where brain tumor patients were imaged before and after gadolinium administration.
Results: A fat-suppressed 3D EPI sequence with a phase encoding bandwidth of around 100 Hz was found to exhibit a good trade-off between geometrical distortions and scan duration. Whole-brain T1 -weighted 3D EPI images with 1.2 mm isotropic voxel size could be acquired in 24 seconds. The WE-rect, its extension, and the SLR RF pulses resulted in reduced magnetization transfer effects and provided a 20% mean increase in gray-white matter contrast.
Conclusion: Using a high phase encoding bandwidth and RF pulses that reduce magnetization transfer effects, a fat-suppressed multi-shot 3D EPI sequence can be used to rapidly acquire isotropic T1 -weighted volumes.
Keywords: 3D; EPI; RF; T1-weighted; binomial; brain.
© 2020 The Authors. Magnetic Resonance in Medicine published by Wiley Periodicals, Inc. on behalf of International Society for Magnetic Resonance in Medicine.
References
-
- Mansfield P. Multi-planar image formation using NMR spin echoes. J Phys C: Solid State Phys. 1977;10:L55-L58.
-
- Belliveau J, Kennedy D, McKinstry R, et al. Functional mapping of the human visual cortex by magnetic resonance imaging. Science. 1991;254:716-719.
-
- Edelman RR, Siewert B, Darby DG, et al. Qualitative mapping of cerebral blood flow and functional localization with echo-planar MR imaging and signal targeting with alternating radio frequency. Radiology. 1994;192:513-520.
-
- Turner R, Le Bihan D, Maier J, Vavrek R, Hedges LK, Pekar J. Echo-planar imaging of intravoxel incoherent motion. Radiology. 1990;177:407-414.
-
- Skare S, Sprenger T, Norbeck O, et al. A 1-minute full brain MR exam using a multicontrast EPI sequence. Magn Reson Med. 2018;79:3045-3054.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical