. 2020 Mar;12(1):22-28.
doi: 10.4055/cios.2020.12.1.22. Epub 2020 Feb 13.

How Many Screws Are Needed for Reliable Stability of Extra-articular Nonosteoporotic Distal Radius Fractures Fixed with Volar Locking Plates?

Affiliations

How Many Screws Are Needed for Reliable Stability of Extra-articular Nonosteoporotic Distal Radius Fractures Fixed with Volar Locking Plates?

Hyoung-Seok Jung et al. Clin Orthop Surg. 2020 Mar.

Abstract

Background: We hypothesized that volar locking pate fixation using a minimum number of screws-four in the distal row and two in the shaft of the plate-will provide sufficient stability for unstable extra-articular fractures of the distal radius. We aimed to compare the biomechanical properties of different numbers and locations of screws in volar locking plate fixation and describe the clinical and radiological outcome of plate fixation using a minimum number of screws for distal radius fractures.

Methods: We divided 48 artificial radius fracture bones into four groups (group A-D) based on the number and location of screws used for fixation with volar locking plates. The artificial bone models were subjected to axial compression and volar bending load with a force of 250 N and 80 N, respectively, for 1,000 cycles at a frequency of 1 Hz. We also retrospectively reviewed 42 patients with unstable, extra-articular, distal radius fractures who were treated with volar locking plate fixation using a minimum number of screws.

Results: Group A (seven distal screws and three proximal screws) had the highest mean stiffness: 303.7 N/mm under axial compression and 61.1 N/mm under volar bending. Compared with group A, group D (four screws in the distal part and two screws in the shaft) showed significantly lower stiffness; therefore, group D was considered inferior in terms of stability. However, in the fatigue test, neither deformation of the metal plate nor detachment or breakage of the metal screws was observed in all groups. In the clinical study, all fractures united without displacement and satisfactory clinical outcome was obtained.

Conclusions: In the dorsally comminuted, extra-articular, nonosteoporotic distal radius fractures, the minimum number of screws-four in the distal row and two in the shaft-in volar locking plate fixation can provide sufficient stability. Further biomechanical studies involving osteoporotic bone will be necessary to confirm the results because volar plate fixation is most commonly used in patients with osteoporosis.

Keywords: Bone screw; Palmar plate; Radius fractures.

PubMed Disclaimer

Conflict of interest statement

CONFLICT OF INTEREST: No potential conflict of interest relevant to this article was reported.

Figures

Fig. 1
Fig. 1. Four different screw configurations in the distal area and shaft of the radius fracture models. (A) Locking screws were placed in all holes in both rows and in the three holes of the shaft. (B) Locking screws were placed in all holes in both rows and in the two holes of the shaft. (C) Locking screws were placed in all holes of the distal row and in the three holes of the shaft. (D) Locking screws were placed in all holes of the distal row and in the two holes of the shaft.
Fig. 2
Fig. 2. Axial compression and volar bending tests. Each specimen was placed in the machine and tested under volar bending (A) and axial compression (B).
Fig. 3
Fig. 3. Preoperative anteroposterior (A) and lateral (B) radiographs of a closed extra-articular distal radius fracture treated with volar locking plate fixation using a minimum number of screws (four in the distal row and two in the proximal shaft). Anteroposterior (C) and lateral (D) radiographs at 12 months after surgery.

References

    1. Mackenney PJ, McQueen MM, Elton R. Prediction of instability in distal radial fractures. J Bone Joint Surg Am. 2006;88(9):1944–1951. - PubMed
    1. Moss DP, Means KR, Jr, Parks BG, Forthman CL. A biomechanical comparison of volar locked plating of intra-articular distal radius fractures: use of 4 versus 7 screws for distal fixation. J Hand Surg Am. 2011;36(12):1907–1911. - PubMed
    1. Souer JS, Lozano-Calderon SA, Ring D. Predictors of wrist function and health status after operative treatment of fractures of the distal radius. J Hand Surg Am. 2008;33(2):157–163. - PubMed
    1. Arora R, Lutz M, Hennerbichler A, Krappinger D, Espen D, Gabl M. Complications following internal fixation of unstable distal radius fracture with a palmar locking-plate. J Orthop Trauma. 2007;21(5):316–322. - PubMed
    1. Lucado AM, Li Z. Static progressive splinting to improve wrist stiffness after distal radius fracture: a prospective, case series study. Physiother Theory Pract. 2009;25(4):297–309. - PubMed