Robotic-Assisted Stereotaxy for Deep Brain Stimulation Lead Implantation in Awake Patients
- PMID: 32147722
- PMCID: PMC8223249
- DOI: 10.1093/ons/opaa029
Robotic-Assisted Stereotaxy for Deep Brain Stimulation Lead Implantation in Awake Patients
Abstract
Background: Robotic-assisted stereotaxy has been increasingly adopted for lead implantation in stereoelectroencephalography based on its efficiency, accuracy, and precision. Despite initially being developed for use in deep brain stimulation (DBS) surgery, adoption for this indication has not been widespread.
Objective: To describe a recent robotic-assisted stereotaxy experience and workflow for DBS lead implantation in awake patients with and without microelectrode recording (MER), including considerations for intraoperative research using electrocorticography (ECoG).
Methods: A retrospective review of 20 consecutive patients who underwent simultaneous bilateral DBS lead implantation using robotic-assisted stereotaxy was performed. Radial error was determined by comparing the preoperative target with the DBS lead position in the targeting plane on postoperative computed tomography. Information regarding any postoperative complications was obtained by chart review.
Results: A novel method for robot coregistration was developed. We describe a standard workflow that allows for MER and/or ECoG research, and a streamlined workflow for cases in which MER is not required. The overall radial error for lead placement across all 20 patients was 1.14 ± 0.11 mm. A significant difference (P = .006) existed between the radial error of the first 10 patients (1.46 ± 0.19 mm) as compared with the second 10 patients (0.86 ± 0.09 mm). No complications were encountered.
Conclusion: Robotic-assisted stereotaxy has the potential to increase precision and reduce human error, compared to traditional frame-based DBS surgery, without negatively impacting patient safety or the ability to perform awake neurophysiology research.
Keywords: Deep brain stimulation; Robotic-assisted stereotaxy.
Copyright © 2020 by the Congress of Neurological Surgeons.
Figures
References
-
- Davies B. A review of robotics in surgery. Proc Inst Mech Eng H. 2000;214(1):129-140. - PubMed
-
- Drake JM, Joy M, Goldenberg A, Kreindler D. Computer- and robot-assisted resection of thalamic astrocytomas in children. Neurosurgery. 1991;29(1):27-33. - PubMed
-
- Kwoh YS, Hou J, Jonckheere EA, Hayati S. A robot with improved absolute positioning accuracy for CT guided stereotactic brain surgery. IEEE Trans Biomed Eng. 1988;35(2):153-160. - PubMed
-
- Glauser D, Fankhauser H, Epitaux M, Hefti JL, Jaccottet A. Neurosurgical robot Minerva: first results and current developments. J Image Guid Surg. 1995;1(5):266-272. - PubMed
-
- Benabid AL, Cinquin P, Lavalle S, Le Bas JF, Demongeot J, de Rougemont J. Computer-driven robot for stereotactic surgery connected to CT scan and magnetic resonance imaging. Technological design and preliminary results. Appl Neurophysiol. 1987;50(1-6):153-154. - PubMed
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources
Miscellaneous