. 2020 Jun;41(5):639-643.
doi: 10.1097/MAO.0000000000002617.

Slim, Modiolar Cochlear Implant Electrode: Melbourne Experience and Comparison With the Contour Perimodiolar Electrode

Affiliations

Slim, Modiolar Cochlear Implant Electrode: Melbourne Experience and Comparison With the Contour Perimodiolar Electrode

Chanan Shaul et al. Otol Neurotol. 2020 Jun.

Abstract

Objective: To describe the tip fold over rate, scalar localization, and speech perception outcomes of the CI532 Slim Modiolar Electrode.

Patients and intervention: All patients receiving the CI532 implant before June 2018.

Main outcome measures: Outcome measures for adults patients include pre- and postoperative speech perception, operative report details, electrode position as determined by X-ray and cone beam computed tomography. Comparison made with previous experience with the Contour perimodiolar electrode (CI512). In the pediatric population tip fold-over rate, measured by intraoperative X-ray, was the exclusive outcome.

Results: One hundred twenty-five CI532 devices were implanted in adults and 69 in children. Electrode tip fold-over occurred in eight adults cases and none among children (4.1%). Cone beam CT scans of 120 out of 125 adult patients confirmed scala tympani (ST) position in all but one case where the electrode had been placed into scala vestibuli. There were no translocations from ST to scala vestibuli. This compares favorably with the CI512 translocation rate of 17%. Speech perception outcomes demonstrated good performance with mean preop phoneme scores of 16.2% (±13) increasing to 64.2% (±14) and 71.6 (±16) 3 and 12-months postop, respectively. Compared with a matched group of CI512 recipients, CI532 recipient phoneme scores were significantly higher 3 and 12-months postop by 4 and 7%, respectively.

Conclusion: The slim modiolar, CI532 electrode has provided very reliable ST position with a low rate of tip fold over. A trend toward better speech perception scores in CI532 compared with CI512 was observed.

PubMed Disclaimer

References

    1. Van Abel KM, Dunn CC, Sladen DP, et al. Hearing preservation among patients undergoing cochlear implantation. Otol Neurotol 2015; 36:416–421.
    1. Wanna GB, Noble JH, Carlson ML, et al. Impact of electrode design and surgical approach on scalar location and cochlear implant outcomes. Laryngoscope 2014; 124: (suppl 6): S1–7.
    1. Briggs RJ, Tykocinski M, Lazsig R, et al. Development and evaluation of the modiolar research array—multi-centre collaborative study in human temporal bones. Cochlear Implants Int 2011; 12:129–139.
    1. Tykocinski M, Saunders E, Cohen LT, et al. The contour electrode array: safety study and initial patient trials of a new perimodiolar design. Otol Neurotol 2001; 22:33–41.
    1. Aschendorff A, Kromeier J, Klenzner T, et al. Quality control after insertion of the nucleus contour and contour advance electrode in adults. Ear Hear 2007; 28:75S–79S.

Publication types