Safety, prosthesis wearing time and health-related quality of life of lower extremity bone-anchored prostheses using a press-fit titanium osseointegration implant: A prospective one-year follow-up cohort study
- PMID: 32150598
- PMCID: PMC7062258
- DOI: 10.1371/journal.pone.0230027
Safety, prosthesis wearing time and health-related quality of life of lower extremity bone-anchored prostheses using a press-fit titanium osseointegration implant: A prospective one-year follow-up cohort study
Abstract
Background: We described safety and functional one-year follow-up outcomes of individuals with lower limb amputation treated with bone-anchored prostheses using titanium press-fit osseointegration implants.
Methods: All consecutive individuals treated between March 2015 and June 2018 with curved osseointegration femur implant (OFI-C) indicated for a long femoral remnant, gamma osseointegration femur implant (OFI-Y) indicated for a short femoral remnant, or osseointegration tibia implant (OTI) were eligible for this study. All adverse events were evaluated, infections were graded as follows: grade 1 and 2: low- and high-grade soft tissue infection, respectively, grade 3: deep bone infection, grade 4: septic implant failure. Functional outcome measures included prosthesis wearing time (PUS), health-related quality of life (GS), and the overall situation as an amputee (GS Q3); evaluated with the Questionnaire of persons with trans-femoral amputation (Q-TFA) before surgery and at one-year follow-up.
Results: Ninety of 91 individuals were included (mean age: 54±14 yrs, 26 females); treated with 53, 16 and 21 OFI-C, OFI-Y and OTI, respectively. Soft tissue infections (grade 1: 11 events, grade 2: 10 events) were treated successfully with antibiotics except in two (OFI-C and OFI-Y), who required additional surgery due to recurrent stoma irritation and peri-stoma abscess drainage. One individual with dysvascular amputation (OTI) developed septic implant loosening and occlusion of the femoral artery resulting in a transfemoral amputation. No aseptic loosening's occurred. One individual (OFI-Y) required stoma surgical refashioning due to soft tissue redundancy. At baseline mean ±SD and median (25th to 75th PCTL) Q-TFA PUS and GS were 52±39, 52(7-90) and 40±19, 42(25-50) and improved significantly to 88±18, 90 (90-100) and 71±15, 75 (67-83) at one-year follow-up. The GS Q3 improved over time.
Conclusion: Titanium osseointegration implants can be safely used within a one-year follow-up period. The performance improved compared to the use of a socket-suspended prosthesis.
Conflict of interest statement
I have read the journal’s policy and the authors of this manuscript have the following competing interests: Henk van de Meent is a shareholder of OTN Implants BV Netherlands. This does not alter our adherence to PLOS ONE policies on sharing data and materials.
Figures








References
-
- Leijendekkers RA, van Hinte G, Frolke JP, van de Meent H, Nijhuis-van der Sanden MW, Staal JB. Comparison of bone-anchored prostheses and socket prostheses for patients with a lower extremity amputation: a systematic review. Disabil Rehabil. 2017;39(11):1045–58. 10.1080/09638288.2016.1186752 . - DOI - PubMed
-
- Branemark R, Branemark PI, Rydevik B, Myers RR. Osseointegration in skeletal reconstruction and rehabilitation: a review. J Rehabil Res Dev. 2001;38(2):175–81. . - PubMed
-
- Leijendekkers RA, van Hinte G, Frolke JP, van de Meent H, Atsma F, Nijhuis-van der Sanden MW, et al. Functional performance and safety of bone-anchored prostheses in persons with a transfemoral or transtibial amputation: a prospective one-year follow-up cohort study. Clin Rehabil. 2019;33(3):450–64. 10.1177/0269215518815215 - DOI - PMC - PubMed
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources