Falsely Elevated Serum Creatinine Associated With IgM Paraproteinemia

Affiliations
Case Reports

Falsely Elevated Serum Creatinine Associated With IgM Paraproteinemia

Hiroyasu Mase et al. Kidney Int Rep. .
No abstract available

PubMed Disclaimer

Figures

Figure 1
Figure 1
Emergence of turbidity, due to mixture of the patient’s serum and the reagents in the kit of our laboratory, and its disappearance following the addition of salt solution. (a) Serum and first reagent; (b) panel (a) and second reagent; (c) panel (b) and 5 mol/l sodium chloride solution.
Figure 2
Figure 2
Reaction curves in creatinine measurement. The curve is distorted in this case (black line), but not in control serum without paraproteinemia with creatinine levels of 5.5 mg/dl (blue line).
Figure 3
Figure 3
Differences in reaction curves depending on sample pretreatment. A sharp elevation due to addition of the second reagent is observed in the serum of this case without pretreatment (black line), which is ablated by elimination of total protein (red line). This distortion is partially attenuated but present after removal of non-IgM protein with 2 agarose-embedded proteins (blue line).
Figure 4
Figure 4
Summary of creatinine measuring methods and mechanisms of interference. Mechanisms involved in this case are highlighted in red. *Adequate pretreatments eliminate interfering substances, including proteins. ID-LC/MS, isotope dilution liquid chromatography/tandem mass spectrometry.

References

    1. Levey A.S., Stevens L.A., Schmid C.H. A new equation to estimate glomerular filtration rate. Ann Intern Med. 2009;150:604–612. - PMC - PubMed
    1. Kidney Disease: Improving Global Outcomes (KDIGO) CKD Work Group KDIGO 2012 clinical practice guideline for the evaluation and management of chronic kidney disease. Kidney Int Suppl. 2013;3:1–150.
    1. Lee E., Collier C.P., White C.A. Interlaboratory variability in plasma creatinine measurement and the relation with estimated glomerular filtration rate and chronic kidney disease diagnosis. Clin J Am Soc Nephrol. 2017;12:29–37. - PMC - PubMed
    1. Killeen A.A., Ashwood E.R., Ventura C.B., Styer P. Recent trends in performance and current state of creatinine assays. Arch Pathol Lab Med. 2013;137:496–502. - PubMed
    1. Drion I., Cobbaert C., Groenier K.H. Clinical evaluation of analytical variations in serum creatinine measurements: why laboratories should abandon Jaffe techniques. BMC Nephrol. 2012;13:133. - PMC - PubMed

Publication types

LinkOut - more resources