Multicenter Study
. 2020 Mar 17;75(10):1111-1124.
doi: 10.1016/j.jacc.2019.12.065.

Early Diagnosis of Myocardial Infarction With Point-of-Care High-Sensitivity Cardiac Troponin I

Collaborators, Affiliations
Free article
Multicenter Study

Early Diagnosis of Myocardial Infarction With Point-of-Care High-Sensitivity Cardiac Troponin I

Jasper Boeddinghaus et al. J Am Coll Cardiol. .
Free article

Erratum in

  • Correction.
    [No authors listed] [No authors listed] J Am Coll Cardiol. 2020 Jun 16;75(23):3001. doi: 10.1016/j.jacc.2020.05.003. J Am Coll Cardiol. 2020. PMID: 32527413 No abstract available.

Abstract

Background: Until now, high-sensitivity cardiac troponin (hs-cTn) assays were mainly developed for large central laboratory platforms.

Objectives: This study aimed to assess the clinical performance of a point-of-care (POC)-hs-cTnI assay in patients with suspected myocardial infarction (MI).

Methods: This study enrolled patients presenting to the emergency department with symptoms suggestive of MI. Two cardiologists centrally adjudicated the final diagnosis using all clinical data including cardiac imaging. The primary objective was to directly compare diagnostic accuracy of POC-hs-cTnI-TriageTrue versus best-validated central laboratory assays. Secondary objectives included the derivation and validation of a POC-hs-cTnI-TriageTrue-specific 0/1-h algorithm.

Results: MI was the adjudicated final diagnosis in 178 of 1,261 patients (14%). The area under the curve (AUC) for POC-hs-cTnI-TriageTrue at presentation was 0.95 (95% confidence interval [CI]: 0.93 to 0.96) and was at least comparable to hs-cTnT-Elecsys (AUC: 0.94; 95% CI: 0.93 to 0.96; p = 0.213) and hs-cTnI-Architect (AUC: 0.92; 95% CI: 0.90 to 0.93; p < 0.001). A single cutoff concentration <3 ng/l at presentation identified 45% of patients at low risk with a negative predictive value (NPV) of 100% (95% CI: 99.4% to 100%). A single cutoff concentration >60 ng/l identified patients at high risk with a positive predictive value (PPV) of 76.8% (95% CI: 68.9% to 83.6%). The 0/1-h algorithm ruled out 55% of patients (NPV: 100%; 95% CI: 98.8% to 100%), and ruled in 18% of patients (PPV: 76.8%; 95% CI: 67.2% to 84.7%). Ruled-out patients had cumulative event rates of 0% at 30 days and 1.6% at 2 years. This study confirmed these findings in a secondary analysis including hs-cTnI-Architect for central adjudication.

Conclusions: The POC-hs-cTnI-TriageTrue assay provides high diagnostic accuracy in patients with suspected MI with a clinical performance that is at least comparable to that of best-validated central laboratory assays. (Advantageous Predictors of Acute Coronary Syndromes Evaluation Study [APACE]; NCT00470587).

Keywords: 0/1-h algorithm; diagnosis of myocardial infarction; guidelines; high-sensitivity cardiac troponin I; point of care; rule in; rule out.