Real-world experience of leadless left ventricular endocardial cardiac resynchronization therapy: A multicenter international registry of the WiSE-CRT pacing system
- PMID: 32165181
- PMCID: PMC7397503
- DOI: 10.1016/j.hrthm.2020.03.002
Real-world experience of leadless left ventricular endocardial cardiac resynchronization therapy: A multicenter international registry of the WiSE-CRT pacing system
Abstract
Background: Biventricular endocardial pacing (BiV ENDO) is a therapy for heart failure patients who cannot receive transvenous epicardial cardiac resynchronization therapy (CRT) or have not responded adequately to CRT. BiV ENDO CRT can be delivered by a new wireless LV ENDO pacing system (WiSE-CRT system; EBR Systems, Sunnyvale, CA), without the requirement for lifelong anticoagulation.
Objective: The purpose of this study was to assess the safety and efficacy of the WiSE-CRT system during real-world clinical use in an international registry.
Methods: Data were prospectively collected from 14 centers implanting the WiSE-CRT system as part of the WiCS-LV Post Market Surveillance Registry. (ClinicalTrials.gov Identifier: NCT02610673).
Results: Ninety patients from 14 European centers underwent implantation with the WiSE-CRT system. Patients were predominantly male, age 68.2 ± 10.5 years, left ventricular ejection fraction 30.6% ± 8.9%, mean QRS duration 180.7 ± 27.0 ms, and 40% with ischemic etiology. Successful implantation and delivery of BiV ENDO pacing was achieved in 94.4% of patients. Acute (<24 hours), 1- to 30-day, and 1- to 6-month complications rates were 4.4%, 18.8%, and 6.7%, respectively. Five deaths (5.6%) occurred within 6 months (3 procedure related). Seventy percent of patients had improvement in heart failure symptoms.
Conclusion: BiV ENDO pacing with the WiSE-CRT system seems to be technically feasible, with a high success rate. Three procedural deaths occurred during the study. Procedural complications mandate adequate operator training and implantation at centers with immediately available cardiothoracic and vascular surgical support.
Keywords: Cardiac resynchronization therapy; Endocardial pacing; Heart failure; Leadless pacing; Nonresponder.
Copyright © 2020 The Authors. Published by Elsevier Inc. All rights reserved.
Figures
References
-
- Petersen S., Rayner M., Wolstenholme J. British Heart Foundation; London: 2002. Coronary Heart Disease Statistics: Heart Failure Supplement.
-
- Nelson G.S., Berger R.D., Fetics B.J. Left ventricular or biventricular pacing improves cardiac function at diminished energy cost in patients with dilated cardiomyopathy and left bundle-branch block. Circulation. 2000;102:3053–3059. - PubMed
-
- Cleland J.G.F., Daubert J.C., Erdmann E. Cardiac Resynchronization-Heart Failure (CARE-HF) Study Investigators. The effect of cardiac resynchronization on morbidity and mortality in heart failure. N Engl J Med. 2005;352:1539–1549. - PubMed
-
- European Society of Cardiology (ESC) European Heart Rhythm Association (EHRA), Brignole M, et al. 2013 ESC guidelines on cardiac pacing and cardiac resynchronization therapy: The Task Force on cardiac pacing and resynchronization therapy of the European Society of Cardiology (ESC). Developed in collaboration with the European Heart Rhythm Association. Europace. 2013;15:1070–1118. - PubMed