Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2021 Dec 1;17(8):e1846-e1850.
doi: 10.1097/PTS.0000000000000665.

Preventing Operating Room Fires: Impact of Surgical Drapes on Oxygen Contamination of the Operative Field

Affiliations

Preventing Operating Room Fires: Impact of Surgical Drapes on Oxygen Contamination of the Operative Field

William C Culp Jr et al. J Patient Saf. .

Abstract

Objective: The aim of the study was to measure underdrape oxygen pooling, surgical site oxygen contamination, and time to restoration of 21% oxygen concentration after cessation of oxygen delivery by measuring oxygen concentration under simulated clinical conditions with various drapes.

Methods: In a 2-part study, oxygen permeability of four differing surgical drapes was measured (Part A) and a mannequin was used to measure underdrape oxygen pooling and surgical site oxygen contamination (Part B). In Part A, a container of high concentration oxygen was sealed with a surgical drape. Oxygen concentrations on both sides of the drape were then measured over time to quantify drape oxygen permeability. Part B included a mannequin model draped for a hypothetical surgical site with oxygen administered by face mask. Oxygen concentration was measured at both the surgical site and under the drape nearest the surgical site.

Results: Oxygen permeability varied significantly between drapes tested. The surgical site oxygen concentration ranged from 20% to 58% (P = 0.0001). The commonly used woven 100% cotton operating room (OR) towel was highly permeable. The plastic occlusive drape created an impermeable barrier, which did not allow for any oxygen contamination but created the longest time to return to 21% oxygen concentration at the underdrape site after cessation of oxygen delivery.

Conclusions: Surgical drapes have varying oxygen permeability and can lead to high concentration underdrape oxygen pooling. Oxygen contamination of the surgical site varies widely based on drape material and may reach dangerously high levels, especially with the cotton OR towel. Surgical drape selection may impact OR fire risk.

PubMed Disclaimer

Conflict of interest statement

W.C.C. reports patent interests in medical safety devices. K.W.M. disclose no conflict of interest.

References

    1. Surgical Fire Prevention. ECRI Institute. Available at: https://www.ecri.org/solutions/accident-investigation-services/surgical-... . Accessed April 7, 2019.
    1. Bruley ME, Arnold TV, Deutsch ES, et al. Surgical fires: decreasing incidence relies on continued prevention efforts. PA Patient Saf Advis . June 2018. Available at: http://patientsafety.pa.gov/ADVISORIES/Pages/201806_SurgicalFires.aspx# . Accessed April 7, 2019.
    1. Hempel S, Maggard-Gibbons M, Nguyen DK, et al. Wrong-site surgery, retained surgical items, and surgical fires: a systematic review of surgical never events. JAMA Surg . 2015;150:796–805.
    1. Apfelbaum JL, Caplan RA, Barker SJ, et al. American Society of Anesthesiologist Task Force on operating room fires. Practice advisory for the prevention and management of operating room fires. Anesthesiology . 2013;118:271–290.
    1. Don’t Fuel the Fire. Anesthesia Patient Safety Foundation. Available at: https://www.apsf.org/wp-content/uploads/collateral/posters/ORFirePrevFly... . Published 2014. Accessed April 7, 2019.

Publication types