Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2020 Mar 17;7(1):18.
doi: 10.1186/s40658-020-0284-5.

Evaluation of a new multipurpose whole-body CzT-based camera: comparison with a dual-head Anger camera and first clinical images

Affiliations

Evaluation of a new multipurpose whole-body CzT-based camera: comparison with a dual-head Anger camera and first clinical images

Cédric Desmonts et al. EJNMMI Phys. .

Abstract

Background: Evaluate the physical performance of the VERITON CzT camera (Spectrum Dynamics, Caesarea, Israel) that benefits from new detection architecture enabling whole-body imaging compared to that of a conventional dual-head Anger camera.

Methods: Different line sources and phantom measurements were performed on each system to evaluate spatial resolution, sensitivity, energy resolution and image quality with acquisition and reconstruction parameters similar to those used in clinical routine. Extrinsic resolution was assessed using 99mTc capillary sources placed successively in air, in a head and in a body phantom filled with background activity. Spectral acquisitions for various radioelements used in nuclear medicine (99mTc, 123I, 201Tl, 111In) were performed to evaluate energy resolution by computing the FWHM of the measured photoelectric peak. Tomographic sensitivity was calculated by recording the total number of counts detected during tomographic acquisition for a set of source geometries representative of different clinical situations. Sensitivity was also evaluated in focus mode for the CzT camera, which consisted of forcing detectors to collect data in a reduced field-of-view. Image quality was assessed with a Jaszczak phantom filled with 350 MBq of 99mTc and scanned on each system with 30-,20-,10- and 5-min acquisition times.

Results: Extrinsic and tomographic resolution in the brain and body phantoms at the centre of the FOV was estimated at 3.55, 7.72 and 6.66 mm for the CzT system and 2.47, 7.75 and 7.72 mm for the conventional system, respectively. The energy resolution measured at 140 keV was 5.46% versus 9.21% for the Anger camera and was higher in a same manner for all energy peaks tested. Tomographic sensitivity for a point source in air was estimated at 236 counts·s-1·MBq-1 and increased to 1159 counts·s-1·MBq-1 using focus mode, which was 1.6 times and 8 times greater than the sensitivity measured on the scintillation camera (144 counts·s-1·MBq-1). Head and body measurements also showed higher sensitivity for the CzT camera in particular with focus mode. The Jaszczak phantom showed high image contrast uniformity and a high signal-to-noise ratio on the CzT system, even when decreasing acquisition time by 6-fold. Representative clinical cases are shown to illustrate these results.

Conclusion: The CzT camera has a superior sensitivity, higher energy resolution and better image contrast than the conventional SPECT camera, whereas spatial resolution remains similar. Introduction of this new technology may change current practices in nuclear medicine such as decreasing acquisition time and activity injected to patient.

Keywords: Molecular imaging; Multipurpose CzT-camera; SPECT; Whole-body imaging.

PubMed Disclaimer

Conflict of interest statement

DA and NA: travel grant (Spectrum dynamics), DA: honourarium (Spectrum dynamics).

The other authors declare that they have no competing interests.

Figures

Fig. 1
Fig. 1
General camera architecture showing the different movements of the 12 detectors (a), schematic principle of the focus mode showing the reduced swipe motion of detectors to a predefined region of interest shown in dark grey (b), design of the detection column consisting of an array of 16 by 128 pixel units (c) and picture of the system (d)
Fig. 2
Fig. 2
Estimated radial spatial resolution of Anger and CzT cameras as a function of the distance to the centre of the field of view (ranging from 0 to 9 cm) measured in air (a) in a head phantom (from IEC Standard 61675-2) (b) and in a body phantom (from IEC Standard 61675-1) (c)
Fig. 3
Fig. 3
Axial reconstructed slices (similar scale with min and max threshold respectively set to 0 and 125%) of the Jaszczak phantom obtained for CzT (a) and Anger cameras (b) centred on cold spheres (top rows) or capillaries (bottom rows) for 30, 20, 10 or 5 min of acquisition time (left to right)
Fig. 4
Fig. 4
Image contrast measured as a function of sphere diameter at different acquisition times for CzT and Anger cameras (a). Total recorded counts (b), integral uniformity (c), and RMS noise (d) calculated for different acquisition times for CzT and Anger cameras
Fig. 5
Fig. 5
Example of a manually drawn VOIs for the thyroid study (a): spherical VOIs of 5 mm of diameter were drawn on lesion corresponding to a cold nodule (blue) and normal tissue (yellow) VOIs were drawn on CzT camera images (left) and copied to Anger camera images (right). The resulting CNRs were computed for each clinical case and both cameras (b)
Fig. 6
Fig. 6
SPECT imaging of 30-min duration performed successively on CzT (with (a) and without (b) CT image fusion) and Anger cameras (c) 48-h after intravenous injection of 5.27 MBq of 223Ra. An uptake on a right shoulder bone metastasis was visible on axial, coronal and sagittal slices (left to right)

Similar articles

Cited by

References

    1. Erlandsson K, Kacperski K, van Gramberg D, Hutton BF. Performance evaluation of D-SPECT: a novel SPECT system for nuclear cardiology. Phys Med Biol. 2009;54(9):2635–2649. doi: 10.1088/0031-9155/54/9/003. - DOI - PubMed
    1. Gambhir SS, Berman DS, Ziffer J, Nagler M, Sandler M, Patton J, et al. A novel high-sensitivity rapid-acquisition single-photon cardiac imaging camera. J Nucl Med. 2009;50(4):635–643. doi: 10.2967/jnumed.108.060020. - DOI - PubMed
    1. Imbert L, Poussier S, Franken PR, Songy B, Verger A, Morel O, et al. Compared performance of high-sensitivity cameras dedicated to myocardial perfusion SPECT: a comprehensive analysis of phantom and human images. J Nucl Med. 2012;53(12):1897–1903. doi: 10.2967/jnumed.112.107417. - DOI - PubMed
    1. Takahashi Y, Miyagawa M, Nishiyama Y, Ishimura H, Mochizuki T. Performance of a semiconductor SPECT system: comparison with a conventional Anger-type SPECT instrument. Ann Nucl Med. 2013;27(1):11–16. doi: 10.1007/s12149-012-0653-9. - DOI - PMC - PubMed
    1. Buechel RR, Herzog BA, Husmann L, Burger IA, Pazhenkottil AP, Treyer V, et al. Ultrafast nuclear myocardial perfusion imaging on a new gamma camera with semiconductor detector technique: first clinical validation. Eur J Nucl Med Mol Imaging. 2010;37(4):773–778. doi: 10.1007/s00259-009-1375-7. - DOI - PubMed

LinkOut - more resources