. 2020 May;30(3):279-285.
doi: 10.1111/vec.12950. Epub 2020 Mar 18.

Evaluation of the Element point-of-care blood gas analyzer for use in horses

Affiliations

Evaluation of the Element point-of-care blood gas analyzer for use in horses

Clarisa R Krueger et al. J Vet Emerg Crit Care (San Antonio). 2020 May.

Abstract

Objective: To compare the Element point-of-care (POC) portable blood gas analyzer with a laboratory-based bench-top reference analyzer using whole blood samples obtained from horses presenting to a referral center with various disorders in order to determine agreement between these analyzers.

Design: Prospective clinical study.

Setting: The study was conducted at a university teaching hospital at moderate altitude.

Animals: One hundred paired samples from 80 horses >1 year of age were collected after obtaining informed client consent. Fifty paired samples were from patients admitted for elective procedures and considered to be healthy, and 50 paired samples were emergency admissions and considered to be critically ill.

Measurements and main results: Paired whole blood samples were evaluated on both the Element POC and Radiometer ABL 800 FLEX analyzers simultaneously, and results were compared. Pearson correlation coefficients between analyzers were calculated. To assess agreement, scatter and Bland-Altman plots were evaluated, and mean difference and 95% limits of agreement were calculated for each analyte. Correlation was either good (0.8-0.92) or excellent (>0.93) for the majority of analytes. All analytes apart from hemoglobin had acceptable agreement, with ≥80% of individual results within agreement targets. Precision targets were acceptable for most analytes, with partial pressure of carbon dioxide (pCO2 ) and calcium (Ca2+ ) exceeding precision targets.

Conclusions: The portable Element POC system had acceptable agreement with the ABL 800 FLEX bench-top analyzer currently in use at the study center when evaluating the majority of analytes from equine whole blood samples.

Keywords: acid-base; critical care monitoring; equine; point-of-care.

PubMed Disclaimer

Similar articles

Cited by

References

REFERENCES

    1. Flatland B, Freeman KP, Vap LM, Harr KE. ASVCP guidelines: quality assurance for point-of-care testing in veterinary medicine. Vet Clin Pathol. 2013;42(4):405-423.
    1. HESKA. Element POC. Available at: http://www.heska.com/products/lab-systems/element-poc-analyzer.aspx?s=7D.... Accessed March 1, 2018.
    1. Bardell D, West E, Mark Senior J. Evaluation of a new handheld point-of-care blood gas analyser using 100 equine blood samples. Vet Anaesth Analg. 2017;44:77-85.
    1. Elmeshreghi TN, Grubb TL, Greene SA, Ragle CA, Wardrop JA. Comparison of enterprise point-of-care and Nova Biomedical Critical Care Xpress analyzers for determination of arterial pH, blood gas, and electrolyte values in canine and equine blood. Vet Clin Pathol. 2018;00:1-10. https://doi.org/10.1111/vcp.12635.
    1. Hackett ES, Traub-Dargatz JL, Knowles JE, Jr, Tarr SF, Dargatz DA. Arterial blood gas parameters of normal foals born at 1500 metres elevation. Equine Vet J. 2010;42(1):59-62.

LinkOut - more resources