Abstract
Aim: The International Continence Society (ICS) has standardized quality control and interpretation of uroflowmetry and urodynamics. We evaluated traces from two large studies of male lower urinary tract symptoms (UPSTREAM and UNBLOCS) against ICS standards of urodynamic equipment and practice.
Methods: Ten percent of uroflowmetry and urodynamics traces were selected at random from hospital sites. A data capture template was designed from the ICS Fundamentals of Urodynamic Practice checklist. Two pretrained blinded assessors extracted the data, with a third assessor to arbitrate. Departmental records of calibration checks and equipment maintenance were scrutinized.
Results: Seven out of twenty-five (28%) departments reported no calibration checks. Four sites (16%) could not provide annual service records. In 32 out of 296 (10.8%) uroflowmetry traces, findings were affected by artifact. One hundred ten urodynamic study traces were reviewed; in 11 records (10%), key pressure traces were incompletely displayed. In 30 (27.2%), reference zero was not set to atmospheric pressure. Resting pressures were outside the expected range for 36 (32.7%). Pressure drift was seen in 18 traces (16.4%). At pressure-flow study commencement, permission to void was omitted in 15 (13.6%). Cough testing after voiding was done in 71.2%, but the resulting cough spikes were significantly different in 16.5%. Erroneous diagnosis of bladder outlet obstruction (BOO) was identified in six cases (5.5%).
Conclusions: Erroneous diagnosis of BOO is a serious error of interpretation, as it could lead to unnecessary surgery. Other errors of standardization, testing, and interpretation were identified with lower risk of adverse implications. Inconsistent documentation of service records mean equipment accuracy is uncertain.
Keywords: LUTS; overactive bladder; standards; urodynamics; uroflowmetry.
© 2020 The Authors. Neurourology and Urodynamics published by Wiley Periodicals LLC.
Comment in
-
Editorial comment on "Quality control of uroflowmetry and urodynamic data from two large multicentre studies of male lower urinary tract symptoms" (NAU-20-0050).Neurourol Urodyn. 2020 Jun;39(5):1615. doi: 10.1002/nau.24379. Epub 2020 Apr 22. Neurourol Urodyn. 2020. PMID: 32320489 No abstract available.
References
REFERENCES
-
- Chua ME, Mendoza J, See M 4th, et al. A critical review of recent clinical practice guidelines on the diagnosis and treatment of non-neurogenic male lower urinary tract symptoms. Can Urol Assoc J. 2015;9:E463-E470.
-
- Gratzke C, Bachmann A, Descazeaud A, et al. EAU guidelines on the assessment of non-neurogenic male lower urinary tract symptoms including benign prostatic obstruction. Eur Urol. 2015;67:1099-1109.
-
- Gammie A, Drake MJ. The fundamentals of uroflowmetry practice, based on International Continence Society good urodynamic practices recommendations. Neurourol Urodyn. 2018;37:S44-S49.
-
- Foster HE, Dahm P, Kohler TS, et al. Surgical management of lower urinary tract symptoms attributed to benign prostatic hyperplasia: AUA Guideline Amendment 2019. J Urol. 2019;202:592-598. https://doi.org/101097JU0000000000000319
-
- Rosier PFWM, Schaefer W, Lose G, et al. International Continence Society Good Urodynamic Practices and Terms 2016: urodynamics, uroflowmetry, cystometry, and pressure-flow study. Neurourol Urodyn. 2017;36:1243-1260.