New approach to technical and clinical evaluation of devices for self-monitoring of blood glucose
- PMID: 3219967
- DOI: 10.2337/diacare.11.8.619
New approach to technical and clinical evaluation of devices for self-monitoring of blood glucose
Abstract
Despite the wide distribution of devices for self-monitoring of blood glucose (SMBG), there is no internationally accepted agreement on a standardized procedure for their evaluation. This is due to incomplete or even inappropriate technical evaluation and to inadequate evaluation criteria for their clinical acceptability. To provide adequate information on the performance of these devices over the whole clinically relevant range for SMBG (30-350 mg/dl), a standardized test procedure has been established for technical (accuracy, precision, and total deviation) and clinical (acceptance analysis) evaluation. To demonstrate the potency of this new approach, the following SMBG devices and test strips were evaluated: Chemstrip bG batch (n = 10), Glucostix batch (n = 2), Accu-Check II (n = 5), Diascan (n = 5), Diascan strip batch (n = 4), Glucocheck SC (n = 4), and Glucometer II (n = 4). The devices and test strips were examined by trained technicians, and in addition, 1 Chemstrip bG batch and 2 Accu-Check II meters were examined by 10 and 5 trained diabetic patients, respectively. Even the best-performing device did not achieve the American Diabetes Association's goal that SMBG measurements should be within 15% of the reference value. Instead, the maximal total deviation within the clinically relevant blood glucose range reached values that equalled deviations from the reference value between 16 and 108%. Three of 36 devices were classified as good, 29 as acceptable, and 4 as unacceptable for clinical use. In conclusion, this new approach to the technical and clinical evaluation of devices for SMBG is easy to perform and provides more realistic and comparable information for clinical use and approval than commonly used methods.
Similar articles
-
Comparative study of home blood glucose monitoring devices: Visidex, Chemstrip bG, Glucometer, and Accu-Chek bG.Diabetes Care. 1983 Nov-Dec;6(6):529-32. doi: 10.2337/diacare.6.6.529. Diabetes Care. 1983. PMID: 6653308
-
Technical and clinical evaluation of an electrochemistry glucose meter: experience in a diabetes center.Diabetes Res Clin Pract. 1998 Oct;42(1):9-15. doi: 10.1016/s0168-8227(98)00077-1. Diabetes Res Clin Pract. 1998. PMID: 9884028
-
Evaluation of two reagent strips and three reflectance meters for rapid determination of blood glucose concentrations.J Vet Intern Med. 1987 Oct-Dec;1(4):170-4. doi: 10.1111/j.1939-1676.1987.tb02011.x. J Vet Intern Med. 1987. PMID: 3506102
-
Analytical Performance Requirements for Systems for Self-Monitoring of Blood Glucose With Focus on System Accuracy: Relevant Differences Among ISO 15197:2003, ISO 15197:2013, and Current FDA Recommendations.J Diabetes Sci Technol. 2015 Jul;9(4):885-94. doi: 10.1177/1932296815580160. Epub 2015 Apr 14. J Diabetes Sci Technol. 2015. PMID: 25872965 Free PMC article. Review.
-
[Standardization and recommendations for the appropriate use of automated devices for self-monitoring blood glucose].Rinsho Byori. 2001 Dec;49(12):1190-8. Rinsho Byori. 2001. PMID: 11797387 Review. Japanese.
Cited by
-
Utility and usability of a dengue NS1 rapid diagnostic as a self-test.Trop Med Health. 2025 Feb 24;53(1):28. doi: 10.1186/s41182-025-00705-9. Trop Med Health. 2025. PMID: 39994768 Free PMC article.
-
SPCE based glucose sensor employing novel thermostable glucose dehydrogenase, FADGDH: blood glucose measurement with 150nL sample in one second.J Diabetes Sci Technol. 2007 Jan;1(1):28-35. doi: 10.1177/193229680700100105. J Diabetes Sci Technol. 2007. PMID: 19888376 Free PMC article.
-
Statistical approach of assessing the reliability of glucose sensors: the GLYCENSIT procedure.J Diabetes Sci Technol. 2008 Nov;2(6):939-47. doi: 10.1177/193229680800200604. J Diabetes Sci Technol. 2008. PMID: 19885283 Free PMC article.
-
Comparison of Accuracy Guidelines for Hospital Glucose Meters.J Diabetes Sci Technol. 2020 May;14(3):546-552. doi: 10.1177/1932296819898277. Epub 2020 Jan 25. J Diabetes Sci Technol. 2020. PMID: 31983225 Free PMC article. Review.
-
Blood glucose reagent strip tests in the operating room: influence of hematocrit, partial pressure of oxygen, and blood glucose level--a comparison of the BM-test 1-44, BM-Accutest, and Satellite G reagent strip systems.J Clin Monit. 1996 Jan;12(1):27-33. doi: 10.1007/BF02025307. J Clin Monit. 1996. PMID: 8732813
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources