. 2020 Mar 16:14:831-835.
doi: 10.2147/OPTH.S246319. eCollection 2020.

A Preliminary Comparative Study of Visual Performance Between Two Newly Commercially Available Monofocal Intraocular Lenses Implanted During Cataract Surgery

Affiliations

A Preliminary Comparative Study of Visual Performance Between Two Newly Commercially Available Monofocal Intraocular Lenses Implanted During Cataract Surgery

Vincenzo Scorcia et al. Clin Ophthalmol. .

Abstract

Purpose: To compare visual outcomes of two newly developed monofocal intraocular lenses (IOLs) (I-Stream H and CT Lucia 211P) implanted in patients undergoing surgery for senile cataract.

Methods: This prospective, comparative, single-center, randomized study included 94 eyes of 94 patients undergoing cataract surgery. Patients were randomized to receive the implantation of I-Stream H IOL (Group 1) or CT Lucia 211P IOL (Group 2). Each patient underwent a complete ocular examination before (V0), 1 month (V1), 3 months (V2) and 6 months (V3) after surgery, including visual acuity testing, objective refraction, slit-lamp biomicroscopy, contrast sensitivity testing and visual function index questionnaire (VF-14).

Results: After surgery, a significant increase of best-corrected visual acuity (BCVA), CS and VF-14 and a significant decrease of SE were detected in both groups at each time point compared to baseline (always p<0.01). No significant differences in the postoperative increase of BCVA (p=0.12), CS (p=0.21) and VF-14 (p=0.31) and decrease of SE (p=0.08) were detected comparing patients from Group 1 and Group 2 at V3. Mean and median prediction error, as well as absolute prediction error values, did not significantly differ between the two groups (p>0.05). Eyes with an actual refraction within 0.25, 0.50 and 1 D were 53.1%, 85.1% and 100% for Group 1 and 55.3%, 87.2% and 100% for Group 2 (p>0.05).

Conclusion: I-Stream H and CT Lucia 211P allowed a satisfied recovery of visual function after senile cataract surgery; both IOLs were shown to have similar outcomes of visual performance.

Keywords: CT Lucia 211P; I-Stream H; IOL; cataract surgery; intraocular lens.

PubMed Disclaimer

Conflict of interest statement

The authors report no funding and no conflicts of interest in this work.

References

    1. Liu YC, Wilkins M, Kim T, Malyugin B, Mehta JS. Cataracts. Lancet. 2017;390(10094):600–612. doi:10.1016/S0140-6736(17)30544-5 - DOI - PubMed
    1. Olson RJ. Cataract surgery from 1918 to the present and future-just imagine! Am J Ophthalmol. 2018;185:10–13. doi:10.1016/j.ajo.2017.08.020 - DOI - PubMed
    1. Wang SY, Stem MS, Oren G, Shtein R, Lichter PR. Patient-centered and visual quality outcomes of premium cataract surgery: a systematic review. Eur J Ophthalmol. 2017;27(4):387–401. doi:10.5301/ejo.5000978 - DOI - PubMed
    1. de Silva SR, Evans JR, Kirthi V, Ziaei M, Leyland M. Multifocal versus monofocal intraocular lenses after cataract extraction. Cochrane Database Syst Rev. 2016;12:CD003169. doi:10.1002/14651858.CD003169.pub4 - DOI - PMC - PubMed
    1. Chylack LT Jr, Wolfe JK, Singer DM, et al. The lens opacities classification system III. The longitudinal study of cataract study group. Arch Ophthalmol. 1993;111(6):831–836. doi:10.1001/archopht.1993.01090060119035 - DOI - PubMed

LinkOut - more resources