. 2020 Jul;47(7):e881-e912.
doi: 10.1002/mp.14157. Epub 2020 May 28.

Principles and applications of multienergy CT: Report of AAPM Task Group 291

Affiliations

Principles and applications of multienergy CT: Report of AAPM Task Group 291

Cynthia H McCollough et al. Med Phys. 2020 Jul.

Erratum in

Abstract

In x-ray computed tomography (CT), materials with different elemental compositions can have identical CT number values, depending on the mass density of each material and the energy of the detected x-ray beam. Differentiating and classifying different tissue types and contrast agents can thus be extremely challenging. In multienergy CT, one or more additional attenuation measurements are obtained at a second, third or more energy. This allows the differentiation of at least two materials. Commercial dual-energy CT systems (only two energy measurements) are now available either using sequential acquisitions of low- and high-tube potential scans, fast tube-potential switching, beam filtration combined with spiral scanning, dual-source, or dual-layer detector approaches. The use of energy-resolving, photon-counting detectors is now being evaluated on research systems. Irrespective of the technological approach to data acquisition, all commercial multienergy CT systems circa 2020 provide dual-energy data. Material decomposition algorithms are then used to identify specific materials according to their effective atomic number and/or to quantitate mass density. These algorithms are applied to either projection or image data. Since 2006, a number of clinical applications have been developed for commercial release, including those that automatically (a) remove the calcium signal from bony anatomy and/or calcified plaque; (b) create iodine concentration maps from contrast-enhanced CT data and/or quantify absolute iodine concentration; (c) create virtual non-contrast-enhanced images from contrast-enhanced scans; (d) identify perfused blood volume in lung parenchyma or the myocardium; and (e) characterize materials according to their elemental compositions, which can allow in vivo differentiation between uric acid and non-uric acid urinary stones or uric acid (gout) or non-uric acid (calcium pyrophosphate) deposits in articulating joints and surrounding tissues. In this report, the underlying physical principles of multienergy CT are reviewed and each of the current technical approaches are described. In addition, current and evolving clinical applications are introduced. Finally, the impact of multienergy CT technology on patient radiation dose is summarized.

Keywords: dual-energy CT; material decomposition; material selective; multienergy CT; virtual monoenergetic; virtual noncontrast.

PubMed Disclaimer

References

REFERENCES

    1. Hounsfield GN. Computerized transverse axial scanning (tomography). 1. Description of system. Br J Radiol. 1973;46:1016-1022.
    1. International Commission on Radiation Units and Measurements. Photon, Electron, Proton and Neutron Interaction Data for Body Tissues (Report 46). Bethesda, MD; 1992.
    1. Hsieh J. Computed Tomography: Principles, Design, Artifacts, and Recent Advances (2nd Edition). Hoboken, NJ: John Wiley & Sons; 2015.
    1. Alvarez RE, Macovski A. Energy-selective reconstructions in x-ray computerized tomography. Phys Med Biol. 1976;21:733-744.
    1. Ergun DL, Mistretta CA, Brown DE, et al. Single-exposure dual-energy computed radiography: improved detection and processing. Radiology. 1990;174:243-249.