Unexpected Shocks From a Subcutaneous Implantable Cardioverter-Defibrillator Despite Attempted Reprogramming and Magnet Use: A Case Report
- PMID: 32224692
- DOI: 10.1213/XAA.0000000000001178
Unexpected Shocks From a Subcutaneous Implantable Cardioverter-Defibrillator Despite Attempted Reprogramming and Magnet Use: A Case Report
Abstract
We present the case of a patient with a subcutaneous implantable cardioverter-defibrillator (S-ICD) in situ. Device interrogation and reprogramming were unsuccessful due to a software mismatch between the device and programmer. The device manufacturer recommended magnet application to suspend antitachycardia therapy. Despite using this strategy, the S-ICD discharged multiple times. The S-ICD has unique perioperative considerations for the anesthesiologist. This case provides an example of the complexity of electrophysiologic devices in current use and the necessity of the anesthesia provider to stay up to date with evolving device management strategies.
References
-
- Weinstock J, Madias C. The subcutaneous defibrillator. Card Electrophysiol Clin. 2017;9:775–783.
-
- van Rees JB, de Bie MK, Thijssen J, Borleffs CJ, Schalij MJ, van Erven L. Implantation-related complications of implantable cardioverter-defibrillators and cardiac resynchronization therapy devices: a systematic review of randomized clinical trials. J Am Coll Cardiol. 2011;58:995–1000.
-
- Friedman DJ, Parzynski CS, Varosy PD, et al. Trends and in-hospital outcomes associated with adoption of the subcutaneous implantable cardioverter defibrillator in the United States. JAMA Cardiol. 2016;1:900–911.
-
- Friedman DJ, Parzynski CS, Heist EK, et al. Ventricular fibrillation conversion testing after implantation of a subcutaneous implantable cardioverter defibrillator: report from the national cardiovascular data registry. Circulation. 2018;137:2463–2477.
-
- Rooke GA, Lombaard SA, Van Norman GA, et al. Initial experience of an anesthesiology-based service for perioperative management of pacemakers and implantable cardioverter defibrillators. Anesthesiology. 2015;123:1024–1032.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical