. 2020 Jun;52(6):474-482.
doi: 10.1055/a-1120-8498. Epub 2020 Mar 30.

A new 12-French plastic stent for unresectable distal malignant biliary obstruction

Affiliations

A new 12-French plastic stent for unresectable distal malignant biliary obstruction

Pierre H Deprez et al. Endoscopy. 2020 Jun.

Abstract

Background: Self-expanding metal stents (SEMSs) are recommended in unresectable distal malignant biliary obstruction. However, problems with dysfunction and migration of these stents are not negligible. We aimed to investigate the effectiveness and safety of a new 12-Fr plastic stent.

Methods: In an observational, prospective study, all consecutive patients who underwent biliary stenting with the 12-Fr stent were considered (index group). Referent groups were a historical cohort, matched by sex, etiology, and metastatic status, including patients with 10-Fr plastic stents and with fully covered and uncovered SEMSs (FCSEMSs and UCSEMSs). Outcomes were stent patency, recurrent biliary obstruction (RBO), technical success, 30-day mortality and adverse events. A post-procedure examination of removed stents was done.

Results: 72 patients (median age 66, range 32 - 94 years, 50 % men) were included (24 index, 48 referents). There were no differences in median stent patency time (P = 0.684). RBO was significantly lower with the 12-Fr compared with the 10-Fr profile stent (50 % vs. 81.3 %, P = 0.04), but no difference was found compared with the FCSEMSs (50 % vs. 43.8 %, P = 0.698). Technical success was 100 %, with no differences in 30-day mortality P = 0.105). The adverse events rate was 4.2 % for both groups (index n = 1, referents n = 2). Of 11 removed 12-Fr plastic stents suspected to be dysfunctional, 7 (64 %) were still patent.

Conclusions: This new 12-Fr plastic stent could be an effective and cheaper alternative to SEMSs in distal malignant biliary obstruction.

PubMed Disclaimer

Conflict of interest statement

This study was supported by Olympus (Berchem, Belgium).