. 2020 Mar 30;11(1):54.
doi: 10.1186/s13244-020-00859-6.

Harmonisation of imaging dosimetry in clinical practice: practical approaches and guidance from the ESR EuroSafe Imaging initiative

Affiliations

Harmonisation of imaging dosimetry in clinical practice: practical approaches and guidance from the ESR EuroSafe Imaging initiative

Eliseo Vano et al. Insights Imaging. .

Abstract

The European Directive 2013/59/EURATOM requires member states of the European Union to ensure justification and optimisation of the radiological procedures and to include information on patient exposure as part of the report of the examinations. The EuroSafe Imaging campaign of the European Society of Radiology created a working group (WG) on "Dosimetry for imaging in clinical practice" with the aim to help with the dosimetry aspects required by European and national regulations. The primary focus topics were selected and a survey among the experts of the WG, allowed suggesting some initial consensus approaches.For information on patient exposure, it was agreed to include the dosimetric values reported by the imaging modalities (validated by a medical physics expert). It was also suggested to prepare educational material on dosimetric quantities for patients. Individual optimisation was considered a challenge, especially for interventional procedures. In these cases, patient and occupational doses should be part of the global optimisation process and trigger levels should be defined to avoid skin radiation injuries. Diagnostic Reference Levels (DRLs) always need to be considered for comparison with periodic patient dose audits. In the case of accidental or unintended exposures, a report should be produced for the Quality Assurance programme, together with an educational note to avoid the repetition of incidents. Dose registry and management systems should allow fulfilling the regulatory requirements of national and European regulations. In a second step, and after the initial experience with the Directive implementation, a wider survey will be considered.

Keywords: Medical imaging; Occupational exposure; Optimisation; Patient exposure; Radiation dose; Regulation.

PubMed Disclaimer

Conflict of interest statement

The authors declare that they have no competing interest.

References

    1. European Council Directive 2013/59/EURATOM on basic safety standards for protection against the dangers arising from exposure to ionising radiation (2014). Off J Eur Union L13;57:1-73 Available at https://eur-lex.europa.eu/LexUriServ/LexUriServ.do?uri=OJ:L:2014:013:000.... Accessed 14 Aug 2019.
    1. European Society of Radiology (ESR) Summary of the European Directive 2013/59/Euratom: Essentials for Health Professionals in Radiology. Insights Imaging. 2015;6:411–417. - PMC - PubMed
    1. Vañó E, Miller DL, Martin CJ et al (2017) ICRP Publication 135: Diagnostic Reference Levels in Medical Imaging. Ann ICRP 46(1):1–143 - PubMed
    1. (2018) Radiation Protection No. 185: European Guidelines on Diagnostic Reference Levels for Paediatric Imaging. Luxembourg: European Commission. 10.2833/486256
    1. European Society of Radiology (ESR) (2018) EuroSafe Imaging Call For Action. Available at http://www.myESR.org/media/659. Accessed 14 Aug 2019.

LinkOut - more resources