A multi-center evaluation of TECHNOSCREEN® ADAMTS-13 activity assay as a screening tool for detecting deficiency of ADAMTS-13
- PMID: 32239643
- DOI: 10.1111/jth.14815
A multi-center evaluation of TECHNOSCREEN® ADAMTS-13 activity assay as a screening tool for detecting deficiency of ADAMTS-13
Abstract
Background: Quantifying A disintegrin-like and metalloprotease with thrombospondin type 1 motif, member 13 (ADAMTS-13) activity enhances thrombotic thrombocytopenic purpura (TTP) diagnosis but most assays are time consuming, technically demanding, and mainly available in reference centers.
Objective: Evaluate a simple, semiquantitative ADAMTS-13 activity screening test for early identification/exclusion of TTP.
Patients/methods: Plasma from 220 patients with suspected thrombotic microangiopathy at three reference centers were tested with TECHNOSCREEN® ADAMTS13 activity screening test in comparison with TECHNOZYM® ADAMTS-13 activity ELISA at two centers, and in-house fluorescence resonance energy transfer assay at the third center. The screening test indicates if ADAMTS-13 activity is at one of four level-indicator points: 0, 0.1, 0.4, or 0.8 IU/mL.
Results: Screen results were interpreted as binary data in that ADAMTS-13 activity was above or below the 0.1 IU/mL TTP clinical threshold. Combining all sites' data, the screen exhibited 88.7% sensitivity, 90.4% specificity, 74.6% positive predictive value, and 96.2% negative predictive value, comparable to published data for quantitative assays. Five samples with quantitative results below the threshold gave screen readings of 0.1 IU/mL and seven marginally above the threshold gave screen readings of zero. All would warrant plasma exchange while the level is quantified. Nine samples with normal/near normal results gave screens of zero and confirmatory quantifications would prompt early treatment withdrawal, as is current practice. One sample generated screen/quantitative results of 0.4/0.00 IU/mL respectively and was the only clear false-negative.
Conclusions: The screening test provides more rapid ADAMTS-13 level evaluation than most currently available assays. Its simple operation renders it suitable for adoption in routine or specialist laboratory environments.
Keywords: ADAMTS-13; ADAMTS-13 activity; screening test; thrombotic microangiopathy; thrombotic thrombocytopenic purpura.
© 2020 International Society on Thrombosis and Haemostasis.
References
REFERENCES
-
- Hassan S, Westwood JP, Ellis D, et al. The utility of ADAMTS13 in differentiating TTP from other acute thrombotic microangiopathies: results from the UK TTP Registry. Br J Haematol. 2015;171:830-835.
-
- Furlan M, Robles R, Lämmle B. Partial purification and characterization of a protease from human plasma cleaving von Willebrand factor to fragments produced by in vivo proteolysis. Blood. 1996;87:4223-4234.
-
- Tsai HM. Physiologic cleavage of von Willebrand factor by a plasma protease is dependent on its conformation and requires calcium ion. Blood. 1996;87:4235-4244.
-
- Joly BS, Coppo P, Veyradier A. Thrombotic thrombocytopenic purpura. Blood. 2017;129:2836-2846.
-
- Scully M, Hunt BJ, Benjamin S, et al. Guidelines on the diagnosis and management of thrombotic thrombocytopenic purpura and other thrombotic microangiopathies. Br J Haematol. 2012;158:323-335.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources