Robustness of a Combined Modified Dixon and PROPELLER Sequence with Two Interleaved Echoes in Clinical Head and Neck MRI
- PMID: 32269186
- PMCID: PMC7952203
- DOI: 10.2463/mrms.mp.2019-0161
Robustness of a Combined Modified Dixon and PROPELLER Sequence with Two Interleaved Echoes in Clinical Head and Neck MRI
Abstract
Purpose: The combination of modified Dixon (mDixon) and periodically rotated overlapping parallel lines with enhanced reconstruction sequence with two interleaved echoes, which promotes uniform fat-suppression and motion insensitivity, has recently become available for commercial magnetic resonance imaging (MRI) scanners. To compare the robustness of this combination sequence with that of standard Cartesian mDixon sequence for fat-suppressed T2-weighted imaging in clinical head and neck MRI.
Methods: Fifty patients with head and neck tumors were involved this study. All patients underwent MRI using both the combination and standard sequences. Two radiologists independently scored motion artifacts and water-fat separation error using a 4-point scale (1, unacceptable; 4, excellent). Furthermore, comprehensive comparative evaluation was performed using a 5-point scale (1, substantially inferior; 5, substantially superior). Data were statistically analyzed using the Wilcoxon signed-rank test.
Results: In the motion artifact assessment, ratings of 3 or 4 points were assigned to 45% (observer-1, 58.0%; observer-2, 32.0%) and 97% (100%; 94.0%) of images for the standard and combination sequences, respectively (P < 0.001). For the water-fat separation error assessment, ratings of 3 or 4 points were assigned to 100% (100%; 100%) and 85% (84.0%; 86.0%) of images, respectively (P < 0.001). In the comprehensive evaluation, of the 100 cases (observer-1, 50; observer-2, 50), 96 were rated at four or five points. In cases with slight or no motion artifacts and water-fat separation errors, the combination sequence was superior to the standard sequence in term of noise and sharpness, and equal in terms of contrast.
Conclusion: Although water-fat separation errors increased significantly in the combination sequence, most of these were acceptable. The significantly decreased motion artifacts in the combination sequence significantly improved image quality overall.
Keywords: Dixon; fat suppression; head and neck; magnetic resonance imaging; periodically rotated overlapping parallel lines with enhanced reconstruction.
Conflict of interest statement
None of the authors has any conflicts of interest regarding this study.
Figures
References
-
- Bolzoni A, Cappiello J, Piazza C, et al. Diagnostic accuracy of magnetic resonance imaging in the assessment of mandibular involvement in oral-oropharyngeal squamous cell carcinoma: a prospective study. Arch Otolaryngol Head Neck Surg 2004; 130:837–843. - PubMed
-
- Stuckensen T, Kovács AF, Adams S, Baum RP. Staging of the neck in patients with oral cavity squamous cell carcinomas: a prospective comparison of PET, ultrasound, CT and MRI. J Craniomaxillofac Surg 2000; 28:319–324. - PubMed
-
- Morimoto Y, Tanaka T, Kito S, et al. Instability of background fat intensity suppression using fat-saturated (FS) MR imaging techniques according to region and reconstruction procedure in patients with oral cancer. Oral Oncol 2004; 40:332–340. - PubMed
-
- Perkins TG, Duijndam A, Eggers H, de Weerdt E, Rijckaert YHE. The next generation fat-free imaging. Available from: https://philipsproductcontent.blob.core.windows.net/assets/20170523/7784... (Accessed: Oct 6, 2019).
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical