Accurate neonatal heart rate monitoring using a new wireless, cap mounted device

Affiliations

Accurate neonatal heart rate monitoring using a new wireless, cap mounted device

Caroline Henry et al. Acta Paediatr. 2021 Jan.

Abstract

Aim: A device for newborn heart rate (HR) monitoring at birth that is compatible with delayed cord clamping and minimises hypothermia risk could have advantages over current approaches. We evaluated a wireless, cap mounted device (fhPPG) for monitoring neonatal HR.

Methods: A total of 52 infants on the neonatal intensive care unit (NICU) and immediately following birth by elective caesarean section (ECS) were recruited. HR was monitored by electrocardiogram (ECG), pulse oximetry (PO) and the fhPPG device. Success rate, accuracy and time to output HR were compared with ECG as the gold standard. Standardised simulated data assessed the fhPPG algorithm accuracy.

Results: Compared to ECG HR, the median bias (and 95% limits of agreement) for the NICU was fhPPG -0.6 (-5.6, 4.9) vs PO -0.3 (-6.3, 6.2) bpm, and ECS phase fhPPG -0.5 (-8.7, 7.7) vs PO -0.1 (-7.6, 7.1) bpm. In both settings, fhPPG and PO correlated with paired ECG HRs (both R2 = 0.89). The fhPPG HR algorithm during simulations demonstrated a near-linear correlation (n = 1266, R2 = 0.99).

Conclusion: Monitoring infants in the NICU and following ECS using a wireless, cap mounted device provides accurate HR measurements. This alternative approach could confer advantages compared with current methods of HR assessment and warrants further evaluation at birth.

Keywords: heart rate; infant; newborn; photoplethysmography; resuscitation; technologies; wireless.

PubMed Disclaimer

References

REFERENCES

    1. Finer N, Rich W. Neonatal resuscitation for the preterm infant: evidence versus practice. J Perinatol. 2010;30(Suppl):S57-66.
    1. Laptook AR, Salhab W, Bhaskar B. Admission temperature of low birth weight infants: predictors and associated morbidities. Pediatrics. 2007;119(3):e643-e649.
    1. de Almeida MF, Guinsburg R, Sancho GA, et al. Hypothermia and early neonatal mortality in preterm infants. J Pediatr. 2014;164(2):271-275.
    1. Wyllie J, Perlman JM, Kattwinkel J, et al. Part 7: neonatal resuscitation: 2015 International Consensus on cardiopulmonary resuscitation and emergency cardiovascular care science with treatment recommendations. Resuscitation. 2015;95:e169-201.
    1. Luong D, Cheung PY, Barrington KJ, et al. Cardiac arrest with pulseless electrical activity rhythm in newborn infants: a case series. Arch Dis Child Fetal Neonatal Ed. 2019;104:F572-F574.

Publication types

LinkOut - more resources