Improved pre-operative diagnostic accuracy for low-grade prosthetic joint infections using second-generation multiplex Polymerase chain reaction on joint fluid aspirate
- PMID: 32296908
- PMCID: PMC7452934
- DOI: 10.1007/s00264-020-04552-7
Improved pre-operative diagnostic accuracy for low-grade prosthetic joint infections using second-generation multiplex Polymerase chain reaction on joint fluid aspirate
Abstract
Background: A major obstacle for the treatment of prosthetic joint infection (PJI) is the identification of the underlying causative organism. While the diagnostic criteria ruling PJI in or out have become ever more accurate, the detection of the causative pathogen(s) still relies mostly on conventional and time-consuming microbial culture. The aim of this study was to evaluate the diagnostic potential of a second-generation multiplex PCR assay (Unyvero ITI G2, Curetis AG, Holzgerlingen, Germany) used on synovial fluid specimens. Our hypothesis was that the method would yield a higher diagnostic accuracy in the pre-operative workup than synovial fluid culture. Thus, a more precise classification of septic and aseptic prosthesis failure could be achieved before revision surgery.
Methods: Prospectively collected frozen joint fluid specimens from 26 patients undergoing arthroplasty revision surgery of the hip or knee were tested as per the manufacturer's protocol. Sensitivities, specificities, positive and negative predictive values as well as positive and negative likelihood ratios with corresponding confidence intervals were estimated using the statistical software R. A combination of the serum C-reactive protein (CRP) level, leukocyte count, erythrocyte sedimentation rate, joint fluid culture, tissue biopsy culture, and tissue biopsy histology served as the gold standard.
Results: Of the 26 patients included in the study, 15 were infected and 11 were aseptic. Conventional joint fluid culture showed a sensitivity of 0.67 and a specificity of 0.91. Joint fluid multiplex PCR yielded a sensitivity of 0.8 and a specificity of 1.0.
Conclusions: Using the second-generation Unyvero ITI cartridge on joint fluid aspirate for the detection of prosthetic joint infection, we were able to achieve a higher diagnostic accuracy than with conventional culture. We conclude that to improve pathogen detection before revision surgery, this method represents a valuable and practicable tool.
Keywords: Agent; Arthrocentesis; Bacteria; Bacterial; Causative organism; DNA; Detection; Diagnosis; PJI; Polymerase chain reaction, diagnostic; Pre-operative; Synovial fluid.
Conflict of interest statement
On behalf of all authors, the corresponding author states that there is no conflict of interest.
Similar articles
-
Diagnosis of prosthetic joint infections using UMD-Universal Kit and the automated multiplex-PCR Unyvero i60 ITI(®) cartridge system: a pilot study.Infection. 2015 Oct;43(5):551-60. doi: 10.1007/s15010-015-0796-4. Epub 2015 May 29. Infection. 2015. PMID: 26021312
-
Synovial fluid multiplex PCR is superior to culture for detection of low-virulent pathogens causing periprosthetic joint infection.Diagn Microbiol Infect Dis. 2018 Feb;90(2):115-119. doi: 10.1016/j.diagmicrobio.2017.10.016. Epub 2017 Nov 2. Diagn Microbiol Infect Dis. 2018. PMID: 29191466
-
Periprosthetic joint infection: Comparison of automated multiplex-PCR Unyvero i60 ITI cartridge system with bacterial culture and real-time PCR.Technol Health Care. 2022;30(4):1005-1015. doi: 10.3233/THC-213248. Technol Health Care. 2022. PMID: 35068428
-
What Markers Best Guide the Timing of Reimplantation in Two-stage Exchange Arthroplasty for PJI? A Systematic Review and Meta-analysis.Clin Orthop Relat Res. 2018 Oct;476(10):1972-1983. doi: 10.1097/01.blo.0000534680.87622.43. Clin Orthop Relat Res. 2018. PMID: 30794241 Free PMC article.
-
Sonication of Arthroplasty Implants Improves Accuracy of Periprosthetic Joint Infection Cultures.Clin Orthop Relat Res. 2017 Jul;475(7):1827-1836. doi: 10.1007/s11999-017-5315-8. Clin Orthop Relat Res. 2017. PMID: 28290115 Free PMC article. Review.
Cited by
-
An update about molecular biology techniques to detect orthopaedic implant-related infections.EFORT Open Rev. 2021 Feb 1;6(2):93-100. doi: 10.1302/2058-5241.6.200118. eCollection 2021 Feb. EFORT Open Rev. 2021. PMID: 33828851 Free PMC article. Review.
-
Diagnostic accuracy of multiplex polymerase chain reaction on tissue biopsies in periprosthetic joint infections.Sci Rep. 2021 Sep 30;11(1):19487. doi: 10.1038/s41598-021-99076-4. Sci Rep. 2021. PMID: 34593961 Free PMC article.
-
Diagnostics in Late Periprosthetic Infections-Challenges and Solutions.Antibiotics (Basel). 2024 Apr 11;13(4):351. doi: 10.3390/antibiotics13040351. Antibiotics (Basel). 2024. PMID: 38667027 Free PMC article. Review.
-
Invited reply to the letter to the editor by McNally et al., 2021.BMC Musculoskelet Disord. 2021 Mar 8;22(1):256. doi: 10.1186/s12891-021-04118-7. BMC Musculoskelet Disord. 2021. PMID: 33685415 Free PMC article. No abstract available.
-
Diagnostic algorithm in septic total knee arthroplasty failure - What is evidence-based?J Orthop. 2021 Jan 28;23:208-215. doi: 10.1016/j.jor.2020.12.020. eCollection 2021 Jan-Feb. J Orthop. 2021. PMID: 33603316 Free PMC article.
References
-
- Parvizi J, Zmistowski B, Berbari EF, Bauer TW, Springer BD, Della Valle CJ, Garvin KL, Mont MA, Wongworawat MD, Zalavras CG. New definition for periprosthetic joint infection: from the workgroup of the musculoskeletal infection society. Clin Orthop Relat Res. 2011;469(11):2992–2994. doi: 10.1007/s11999-011-2102-9. - DOI - PMC - PubMed
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical
Research Materials
Miscellaneous