Efficacy and safety of palliative endobiliary radiofrequency ablation using a novel temperature-controlled catheter for malignant biliary stricture: a single-center prospective randomized phase II TRIAL
- PMID: 32488654
- DOI: 10.1007/s00464-020-07689-z
Efficacy and safety of palliative endobiliary radiofrequency ablation using a novel temperature-controlled catheter for malignant biliary stricture: a single-center prospective randomized phase II TRIAL
Abstract
Background: Endobiliary radiofrequency ablation (EB-RFA) has emerged as a palliative treatment for malignant biliary strictures (MBSs); however, concerns about complications related to thermal injury remain. In this study, we evaluated the efficacy and safety of EB-RFA with a novel catheter for MBS.
Methods: Patients with inoperable cancer causing MBS were randomly assigned to either the radiofrequency ablation (RFA) group or the non-RFA group. The RFA group underwent EB-RFA at the stricture site with a temperature-controlled catheter (ELRA™; STARmed Co., Goyang, Korea) followed by deployment of a self-expanding metal stent (SEMS). For the non-RFA group, only SEMS placement was performed. The duration of stent patency, overall survival (OS), and 30-day complication rate were evaluated. This trial was registered at ClinicalTrials.gov (number NCT02646514).
Results: A total of 48 patients were enrolled (24 in each group). During a median follow-up period of 135.0 days (RFA group) and 119.5 days (non-RFA group), the 90-day stent patency rate, median duration of stent patency, and median OS were not different between the groups (58.3% vs. 45.8% [P = 0.386], 132.0 days vs. 116.0 days [P = 0.440], and 244.0 days vs. 180.0 days [P = 0.281], respectively). In the RFA group, procedure-related complications including thermal injury-related complications, such as bile duct perforation or hemobilia, were not reported. The early complication (< 7 days) rates were not different between the groups (4.2% vs. 12.5%, P = 0.609), and there were no late complications (7-30 days) in both groups.
Conclusion: EB-RFA with a temperature-controlled catheter followed by SEMS placement for patients with inoperable MBS can be safe and feasible with acceptable biliary patency.
Keywords: Biliary stricture; Endobiliary; Malignant; Metal stent; Palliative; Radiofrequency ablation.
References
-
- Abraham NS, Barkun JS, Barkun AN (2002) Palliation of malignant biliary obstruction: a prospective trial examining impact on quality of life. Gastrointest Endosc 56(6):835–841 - DOI
-
- Rustagi T, Jamidar PA (2015) Endoscopic treatment of malignant biliary strictures. Curr Gastroenterol Rep 17(1):426 - PubMed
-
- Sawas T, Al Halabi S, Parsi MA, Vargo JJ (2015) Self-expandable metal stents versus plastic stents for malignant biliary obstruction: a meta-analysis. Gastrointest Endosc 82(2):256–267.e257 - DOI
-
- Davids PH, Groen AK, Rauws EA, Tytgat GN, Huibregtse K (1992) Randomised trial of self-expanding metal stents versus polyethylene stents for distal malignant biliary obstruction. Lancet 340(8834–8835):1488–1492 - DOI
-
- Cho JH, Jeong S, Kim EJ, Kim JM, Kim YS, Lee DH (2018) Long-term results of temperature-controlled endobiliary radiofrequency ablation in a normal swine model. Gastrointest Endosc 87(4):1147–1150 - DOI