Effects of different levels of CT iterative reconstruction on low-contrast detectability and radiation dose in patients of different sizes: an anthropomorphic phantom study
- PMID: 32495272
- DOI: 10.1007/s11547-020-01228-5
Effects of different levels of CT iterative reconstruction on low-contrast detectability and radiation dose in patients of different sizes: an anthropomorphic phantom study
Abstract
Purpose: The purpose of this study was to verify the maintenance of low-contrast detectability at different CT dose reduction levels, in patients of different sizes, as a consequence of the application of iterative reconstruction at different strengths combined with tube current modulation.
Methods: Anthropomorphic abdominal phantoms of two sizes (small and large) were imaged at a fixed noise with iterative algorithm ASIR-V percentages in the range between 0 and 70% and corresponding dose reductions in the range of 0-83%. A total of 1400 images with and without liver low-contrast simulated lesions were evaluated by five radiologists, using the receiver operating characteristics (ROC) paradigm and evaluating the area under the ROC curve (AUC). The human observer results were then compared with AUC obtained with a channelized Hotelling observer (CHO). CNR values were also calculated.
Results: For the small phantom, the AUC values lie between 0.90 and 0.93 for human evaluations of images acquired without iterative reconstruction, with 30% ASIR-V and with 50% ASIR-V. The AUC decreased significantly to 0.81 (p = 0.0001) at 70% ASIR-V. The CHO results were in coherence with human observer scores. Also, similar results were observed for the large size phantom. CNR values were stable for the different ASIR-V percentages.
Conclusions: The iterative algorithm maintained the low-contrast detectability up to a dose reduction of about 70%, following application of a 50% ASIR-V combined with automatic tube current modulation, regardless of the phantom size. At further dose reductions using greater iterative percentages, a significant decrease in detectability was observed.
Keywords: Computed tomography; Iterative reconstruction; Low-contrast detectability; Model observer; Patient dose reduction.
References
-
- Beister M, Kolditz D, Kalender WA (2012) Iterative reconstruction methods in X-ray CT. Phys Med 28(2):94–108 - DOI
-
- Kordolaimi SD, Argentos S, Pantos I, Kelekis NL, Efstathopoulos EP (2013) A new era in computed tomographic dose optimization: the impact of iterative reconstruction on image quality and radiation dose. J Comput Assist Tomogr 37(6):924–931 - DOI
-
- Schindera ST, Odedra D, Raza SA, Kim TK, Jang HJ, Szucs-Farkas Z, Rogalla P (2013) Iterative reconstruction algorithm for CT: Can radiation dose be decreased while low-contrast detectability is preserved? Radiology 269(2):511–518 - DOI
-
- Goenka AH, Herts BR, Obuchowski NA, Primak AN, Dong F, Karim W, Baker ME (2014) Effect of reduced radiation exposure and iterative reconstruction on detection of low-contrast low-attenuation lesions in an anthropomorphic liver phantom: an 18-reader study. Radiology 272(1):154–163 - DOI
-
- Greffier J, Hamard A, Pereira F et al (2020) Image quality and dose reduction opportunity of deep learning image reconstruction algorithm for CT: a phantom study. Eur Radiol. https://doi.org/10.1007/s00330-020-06724-w - DOI - PubMed
MeSH terms
LinkOut - more resources
Full Text Sources
Medical