Durability of the St. Jude Epic Supra Bioprosthetic Valve in the Pulmonary Position
- PMID: 32505797
- DOI: 10.1053/j.semtcvs.2020.05.026
Durability of the St. Jude Epic Supra Bioprosthetic Valve in the Pulmonary Position
Abstract
Epic Supra valves have been used off-label in the pulmonary position. We aim to evaluate the durability of Epic valves in the pulmonary position. We performed a retrospective review of all Epic valves placed in the pulmonary position from October 2008 to May 2019. Time-to-event analysis was performed using Kaplan-Meier estimates to evaluate freedom from valve intervention, moderate pulmonary regurgitation, and peak velocity greater than 3.5 m/s. Valve dysfunction was a composite of all 3 end points. A total of 79 patients had Epic valves implanted in the pulmonary position. Median age was 18.5 years (15th-85th percentile 11.2-41.0). In total, 1 (1%) 19 mm valve, 4 (5%) 21 mm valves, 8 (10%) 23 mm valves, 23 (29%) 25 mm valves, and 43 (54%) 27 mm valves were implanted. There were no deaths or transplants. Median follow-up was 3.1 years (interquartile range 1.0-5.5). At 5 years, freedom from valve intervention was 95%, freedom from valve dysfunction was 68%, freedom from moderate pulmonary regurgitation was 73%, and freedom from peak velocity greater than 3.5 m/s was 82%. Epic Supra valves provide an acceptable valve replacement in the pulmonary position for children and adults. Longer follow-up is needed to determine valve durability through the entirety of the valve life expectancy.
Keywords: Bioprosthetic valve; Epic supra; Pulmonary valve.
Copyright © 2020 Elsevier Inc. All rights reserved.
Comment in
-
Commentary: Epic Supra Valve Just as Good for Pulmonary Valve Replacement - But is it Good Enough?Semin Thorac Cardiovasc Surg. 2021 Spring;33(1):193-194. doi: 10.1053/j.semtcvs.2020.07.011. Epub 2020 Aug 25. Semin Thorac Cardiovasc Surg. 2021. PMID: 32858215 No abstract available.
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources