. 2020 Aug:129:104446.
doi: 10.1016/j.jcv.2020.104446. Epub 2020 May 21.

Evaluation of rapid diagnosis of novel coronavirus disease (COVID-19) using loop-mediated isothermal amplification

Affiliations

Evaluation of rapid diagnosis of novel coronavirus disease (COVID-19) using loop-mediated isothermal amplification

Yutaro Kitagawa et al. J Clin Virol. 2020 Aug.

Abstract

With the rapid spread of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), there is an urgent need for more rapid and simple detection technologies at the forefront of medical care worldwide. In this study, we evaluated the effectiveness of the Loopamp® 2019-SARSCoV-2 Detection Reagent Kit, which uses loop-mediated isothermal amplification (LAMP) technology. In this protocol, cDNA is synthesized from SARS-CoV-2 RNA using reverse transcriptase, followed by DNA amplification under isothermal conditions in one step. The RT-LAMP test kit amplified the targeted RNA of a SARS-CoV-2 isolate with a detection limit of 1.0 × 101 copies/μL, which was comparable to the detection sensitivity of quantitative reverse transcription PCR (RT-qPCR). Comparison with the results of RT-qPCR for 76 nasopharyngeal swab samples from patients with suspected COVID-19 showed a sensitivity of 100 % and a specificity of 97.6 %. In the 24 RNA specimens derived from febrile Japanese patients with or without influenza A, no amplification was observed using RT-LAMP. RT-LAMP could be a simple and easy-to-use diagnostic tool for the detection of SARS-CoV-2.

Keywords: COVID-19; LAMP; RT-qPCR; SARS-CoV-2.

PubMed Disclaimer

Figures

Fig. 1
Fig. 1
Detection limit of the RT-LAMP method for SARS-CoV-2 with 10-fold serial dilutions of standard RNA.
Fig. 2
Fig. 2
Visual confirmation of turbidity under natural light after completion of the LAMP reaction. Photographs of microtubes after the end of the RT-LAMP reaction under natural light. N, negative control; P, positive control.

Similar articles

Cited by

References

    1. Huang C., Wang Y., Li X. Clinical features of patients infected with 2019 novel coronavirus in Wuhan, China. Lancet. 2020;395:497–506. - PMC - PubMed
    1. Li Q., Guan X., Wu P. Early transmission dynamics in Wuhan, China, of novel coronavirus-infected pneumonia. N. Engl. J. Med. 2020 doi: 10.1056/NEJMoa2001316. In Press. - DOI - PMC - PubMed
    1. Wang J., Zhou M., Liu F. Exploring the reasons for healthcare workers infected with novel coronavirus disease 2019 (COVID-19) in China. J. Hosp. Infect. 2020 doi: 10.1016/j.jhin.2020.03.002. In Press. - DOI - PMC - PubMed
    1. Boehme C.C., Nabeta P., Henastroza G. Operational feasibility of using loop mediated isothermal amplification (LAMP) for the diagnosis of pulmonary TB in microscopy centers of developing countries. J. Clin. Microbiol. 2007;45:1936–1940. - PMC - PubMed
    1. Mitarai S., Okumura M., Toyota E. Evaluation of a simple loop mediated isothermal amplification test kit for the diagnosis of tuberculosis. Int. J. Tuberc. Lung Dis. 2011;l5:1211–1217. - PubMed

Publication types

Supplementary concepts