Comparative Study
. 2020 Aug:129:104427.
doi: 10.1016/j.jcv.2020.104427. Epub 2020 May 8.

Comparison of a laboratory-developed test targeting the envelope gene with three nucleic acid amplification tests for detection of SARS-CoV-2

Affiliations
Comparative Study

Comparison of a laboratory-developed test targeting the envelope gene with three nucleic acid amplification tests for detection of SARS-CoV-2

Philip L Bulterys et al. J Clin Virol. 2020 Aug.

Abstract

Background: Numerous nucleic acid amplification tests, including real-time, reverse transcription PCR (rRT-PCR) and isothermal amplification methods, have been developed to detect SARS-CoV-2 RNA, including many that have received emergency use authorization (EUA). There is a need to assess their test performance relative to one another.

Objectives: The aim of this study was to compare the test performance of a high complexity laboratory-developed rRT-PCR EUA from Stanford Health Care (SHC) targeting the SARS-CoV-2 envelope (E) gene with other tests: the Atila isothermal amplification assay targeting the nucleocapsid (N) gene and open reading frame 1ab (ORF1ab), the Altona E and spike (S) multiplex, real-time RT-PCR, and the US Centers for Disease Control and Prevention (CDC) N1 and N2 rRT-PCRs.

Study design: A diagnostic comparison study was performed by testing nasopharyngeal samples from persons under investigation for coronavirus disease 2019 (COVID-19). Assay performance was assessed by percent agreement and Cohen's kappa coefficient.

Results: Positive percent agreement with the SHC EUA reference assay was 82.8 % (95 % confidence interval (CI) 65.0 to 92.9) for Atila, 86.7 % (95 % CI 69.7 to 95.3) for the Altona E and S targets, and 86.7 % (95 % CI 69.7 to 95.3) and 90.0 % (95 % CI 73.6 to 97.3), for the CDC N1 and N2 targets, respectively. All assays demonstrated 100 % negative percent agreement. Kappa coefficients ranged from 0.86 to 0.92, indicating excellent agreement.

Conclusions: Performance was comparable among the SARS-CoV-2 nucleic acid amplification methods tested, with a limited number of discrepancies observed in specimens with low viral loads.

Keywords: COVID-19; Coronavirus; Diagnostics; SARS-CoV-2.

PubMed Disclaimer

Conflict of interest statement

Declaration of Competing Interest The authors declare no conflicts of interest.

Similar articles

Cited by

References

    1. Corman V.M., Landt O., Kaiser M. Detection of 2019 novel coronavirus (2019-nCoV) by real-time RT-PCR. Euro Surveill. 2020;25(3) - PMC - PubMed
    1. Waggoner J.J., Stittleburg V., Pond R. Triplex real-time RT-PCR for severe acute respiratory syndrome coronavirus 2. Emerg Infect Dis. 2020;26(7) - PMC - PubMed
    1. Nalla A.K., Casto A.M., Huang M.W. Comparative performance of SARS-CoV-2 detection assays using seven different Primer/Probe sets and one assay kit. J. Clin. Microbiol. 2020 - PMC - PubMed
    1. Hogan C.A., Sahoo M.K., Huang C. Comparison of the panther fusion and a laboratory-developed test targeting the envelope gene for detection of SARS-CoV-2. J. Clin. Virol. 2020 - PMC - PubMed
    1. U.S. Food and Drug Administration . 2020. Stanford Health Care Clinical Virology Laboratory SARS-CoV-2 Test EUA Summary.https://www.fda.gov/media/136818/download Published. Accessed April 10 2020.