Multicenter Study
. 2021 Jul 1;98(1):45-54.
doi: 10.1002/ccd.29065. Epub 2020 Jun 17.

Evaluation of the safety and efficacy of the Cobra PzF NanoCoated coronary stent in routine, consecutive, prospective, and high-risk patients: The e-Cobra study

Affiliations
Multicenter Study

Evaluation of the safety and efficacy of the Cobra PzF NanoCoated coronary stent in routine, consecutive, prospective, and high-risk patients: The e-Cobra study

Luc Maillard et al. Catheter Cardiovasc Interv. .

Abstract

Background: The Cobra PzF coronary stent is cobalt chromium with flat thin struts, nano-coated with Polyzene-F that enhance rapid reendothelialization and reduce the risk of stent thrombosis and restenosis. It is designed to overcome shortfalls of BMS and DES in patients requiring short DAPT duration.

Aims: To report procedural and 1-year clinical outcomes following Cobra PzF stent implantation in routine practice PCI.

Methods: e-Cobra registry is a multicenter prospective study to evaluate Cobra PzF stent in routine practice in patients deemed appropriate for short DAPT after PCI. The primary endpoint was MACE rate at 12 months (Cardiac death, MI, TLR). The secondary endpoint was definite stent thrombosis at 12 months.

Results: Among 940 patients (72% men, 72.8 ± 13.4 years) with multiple co-morbidities, 47% had acute coronary syndromes, and 62% were defined as high bleeding risk. A total of 1,229 lesions were treated with 1,314 stents. 36% of patients had lesion type B2 or C classification. Angiographic success was achieved in all cases. One-year follow-up was available for 97% of patients. The primary endpoint occurred in 9.0% of patients, including cardiac death 3.7%, MI 4.8%, and TLR 4.3%. Definite stent thrombosis occurred in six out of 915 (0.7%).

Conclusion: The Cobra PzF stent was safe and effective in routine practice patients and seems feasible in situations when short DAPT or Mono Antiplatelet Therapy (MAPT) is needed. One-year follow-up was associated with satisfactory clinical outcomes and validate previously reported data.

Keywords: Cobra PzF stent; Polyzene-F; stent.

PubMed Disclaimer

Comment in

References

REFERENCES

    1. Caixeta A, Leon MB, Lansky AJ, et al. 5-year clinical outcomes after sirolimus-eluting stent implantation insights from a patient-level pooled analysis of 4 randomized trials comparing sirolimus-eluting stents with bare-metal stents. J Am Coll Cardiol. 2009;54(10):894-902.
    1. Lee MS, Wolfe M, Stone GW. Transradial versus transfemoral percutaneous coronary intervention in acute coronary syndromes: re-evaluation of the current body of evidence. JACC Cardiovasc Interv. 2013;6(11):1149-1152.
    1. Valgimigli M. The ESC DAPT guidelines 2017. Eur Heart J. 2018;39(3):187-188.
    1. Bittl JA, Baber U, Bradley SM, Wijeysundera DN. Duration of dual antiplatelet therapy: a systematic review for the 2016 ACC/AHA guideline focused update on duration of dual antiplatelet therapy in patients with coronary artery disease. A Report of the American College of Cardiology/American Heart Association Task Force on Clinical Practice Guidelines. J Am Coll Cardiol. 2016;68(10):1116-1139.
    1. Morice MC, Urban P, Greene S, Schuler G, Chevalier B. Why are we still using coronary bare-metal stents? J Am Coll Cardiol. 2013;61(10):1122-1123.

Publication types

Substances

Grants and funding