. 2021 Mar;197(3):198-208.
doi: 10.1007/s00066-020-01651-7. Epub 2020 Jun 19.

A randomized in vitro evaluation of transient and permanent cardiac implantable electronic device malfunctions following direct exposure up to 10 Gy

Affiliations

A randomized in vitro evaluation of transient and permanent cardiac implantable electronic device malfunctions following direct exposure up to 10 Gy

Maria Daniela Falco et al. Strahlenther Onkol. 2021 Mar.

Abstract

Background and purpose: High-dose 6‑MV radiotherapy may cause cardiac implantable electronic devices (CIEDs) to malfunction. To assess CIED malfunctions resulting from direct exposure up to 10 Gy, 100 pacemakers (PMs) and 40 implantable cardioverter-defibrillators (ICDs) were evaluated.

Materials and methods: CIEDs underwent baseline interrogation. In ICDs, antitachycardia therapies were disabled via the programmer while the detection windows were left enabled. A computed tomography (CT) scan was performed to build the corresponding treatment plan. CIEDs were "blinded" and randomized to receive single doses of either 2, 5, or, 10 Gy via a 6-MV linear accelerator (linac) in a water phantom. Twenty-two wireless telemetry-enabled CIEDs underwent a real-time session, and their function was recorded by the video camera in the bunker. The CIEDs were interrogated after exposure and once monthly for 6 months.

Results: During exposure, regardless of dose, 90.9% of the CIEDs recorded electromagnetic interference, with 6 ICDs (27.3%) reporting pacing inhibition and inappropriate arrhythmia detections. After exposure, a backup reset was observed in 1 PM (0.7% overall, 1% among PMs), while 7 PMs (5% overall, 7% among PMs) reported battery issues (overall immediate malfunction rate was 5.7%). During follow-up, 4 PMs (2.9% overall; 4% among PMs) and 1 ICD (0.7% overall; 2.5% among ICDs) reported abnormal battery depletion, and 1 PM (0.7% overall; 1% among PMs) reported a backup reset (overall late malfunction rate was 4.3%).

Conclusion: Apart from transient electromagnetic interference, last-generation CIEDs withstood direct 6‑MV exposure up to 10 Gy. Permanent battery or software errors occurred immediately or later only in less recent CIEDs.

Keywords: Arrhythmias; Electromagnetic interference; Implantable cardioverter-defibrillator; Pacemaker; Radiotherapy.

PubMed Disclaimer

Similar articles

Cited by

References

    1. Mond HG, Proclemer A (2011) The 11th world survey of cardiac pacing and implantable cardioverter-defibrillators: calendar year 2009—a world society of arrhythmia’s project. Pacing Clin Electrophysiol 34:1013–1027. https://doi.org/10.1111/j.1540-8159.2011.03150.x - DOI - PubMed
    1. Køber L, Thune JJ, Nielsen CJ, Haarbo J, Videbæk L, Korup E, Jensen G, Hildebrandt P, Steffensen FH, Bruun NE, Eiskjær H, Brandes A, Thøgersen AM, Gustafsson F, Egstrup K, Videbæk R, Hassager C, Svendsen JH, Høfsten DE, Torp-Pedersen C, Pehrson S, DANISH Investigators (2016) Defibrillator implantation in patients with nonischemic systolic heart failure. N Engl J Med 375(13):1221–1230. https://doi.org/10.1056/NEJMoa1608029 - DOI - PubMed
    1. Jukema WJ, Timal RJ, Rotmans JI, Hensen LCR, Buiten MS, de Bie MK, Putter H, Zwinderman AH, van Erven L, Krol-van Straaten MJ, Hommes N, Gabreëls B, van Dorp W, van Dam B, Herzog CA, Schalij MJ, Rabelink TJ, ICD2 Trial Investigators (2019) Prophylactic of Implantable cardioverter-defibrillators in the prevention of sudden cardiac death in dialysis patients. Circulation 139(23):2628–2638. https://doi.org/10.1161/circulationaha.119.039818 - DOI - PubMed
    1. Jemal A, Bray F, Center MM, Ferlay J, Ward E, Forman D (2011) Global cancer statistics. CA Cancer J Clin 61:69–90. https://doi.org/10.3322/caac.20107 - DOI - PubMed
    1. Passman R, Kadish A (2007) Sudden death prevention with implantable devices. Circulation 116:561–571. https://doi.org/10.1161/circulationaha.106.655704 - DOI - PubMed

MeSH terms