Technical Delivery Parameters of 2000 Proton Treatment Courses
- PMID: 32582812
- PMCID: PMC7038916
- DOI: 10.14338/IJPT-19-00066.1
Technical Delivery Parameters of 2000 Proton Treatment Courses
Abstract
Purpose: To summarize the technical delivery parameters of proton plans delivered at the Mayo Clinic in Rochester, Minnesota.
Materials and methods: The database of treated patient proton plans was queried to extract field parameters such as gantry angle, patient support angle, minimum and maximum water-equivalent depth (WED) treated, number of layers, field size, patient orientation, and monitor units. The plans were analyzed in aggregate, by disease site, and by fractionation.
Results: There were 2963 proton plans for 2023 distinct treatment sites delivered between June 2015 and September 2018. The mean number of fields per plan was 2.8. The mean number of energy layers per field was 51.9. The mean monitor unit per field was 117.4. The median maximum field dimension was 12.4 cm; 95% of the fields had a maximum dimension < 28.7 cm, and the maximum field dimension was 39.8 cm. The median maximum field WED was 16.4 cm; 95% of the fields reached a maximum WED of ≤ 26.4 cm, and the maximum field WED was 32.4 cm.
Conclusion: A large variety of disease sites were treated using the maximum field size (40 cm) and WED (32.4 cm) capabilities of our half-gantry system.
Keywords: gantry; pencil-beam scanning; proton therapy.
© Copyright 2020 The Author(s).
Conflict of interest statement
Conflicts of Interest: Anita Mahajan, MD, is an Associate Editor of the International Journal of Particle Therapy. The authors have no other relevant conflicts of interest to disclose.
Figures
References
-
- Particle Therapy Co-operative Group. Facilities in operation. 2019 Apr; https://www.ptcog.ch/index.php/facilities-in-operation Updated. Accessed May 1, 2019.
-
- Yan S, Lu HM, Flanz J, Adams J, Trofimov A, Bortfeld T. Reassessment of the necessity of the proton gantry: analysis of beam orientations from 4332 treatments at the Massachusetts General Hospital proton center over the past 10 years. Int J Radiat Oncol Biol Phys. 2016;95:224–33. - PubMed
-
- Courneyea L, Beltran C, Tseung HS, Yu J, Herman MG. Optimizing mini-ridge filter thickness to reduce proton treatment times in a spot-scanning synchrotron system. Med Phys. 2014;41:061713. - PubMed
LinkOut - more resources
Full Text Sources