Improved real-time recordings using the fourth-generation cryoballoon technology-detection of dual fascicle electrograms
- PMID: 32591926
- DOI: 10.1007/s10840-020-00809-8
Improved real-time recordings using the fourth-generation cryoballoon technology-detection of dual fascicle electrograms
Abstract
Background: We aimed to analyze the rate of visualization of real-time (RT) recordings and dual fascicle electrograms in our first series of patients treated with the fourth-generation cryoballoon (CB) device.
Methods: All consecutive patients who underwent CB ablation using the fourth-generation technology were included in the analysis. In all procedures, we used a 28-mm CB placed via a single transseptal puncture guided by intracardiac ultrasound. A 20-mm octapolar intraluminal circular catheter was used for intracardiac recordings. A single 180-s freeze strategy was employed.
Results: A total of 129 patients (72.9% male, mean age 60.1 ± 10.9 years) were enrolled in the study. RT recordings were detected in 445 (86.2%) pulmonary veins (PVs). Specifically, RT recordings were visualized in 115 left superior PVs (89.2%), 107 left inferior PVs (82.9%), 118 right superior PVs (91.4%), and 105 right inferior PVs (81.3%). Dual fascicle electrograms could be observed only in patients in sinus rhythm, in 23 of 396 PVs (5.8%). They were most commonly observed in the left superior PV (8.1%), somewhat less in both inferior veins (6.1%), while this phenomenon was least frequent in the right superior PV (3.0%).
Conclusions: By using the fourth-generation CB, we report a specific pattern of isolation represented by a sequential isolation of two apparently distinct PV fascicles during a single CB freeze delivery. This phenomenon occurred in 5.8% of PV ablations. Of note, the rate of visualization of RT isolation with this novel CB was very high and could be documented in 86.2% of PVs.
Keywords: Atrial fibrillation; Catheter ablation; Cryoballoon; Real-time recordings.
© 2020. Springer Science+Business Media, LLC, part of Springer Nature.
References
-
- Calkins H, Hindricks G, Cappato R, Kim YH, Saad EB, Aguinaga L, et al. 2017 HRS/EHRA/ECAS/APHRS/SOLAECE expert consensus statement on catheter and surgical ablation of atrial fibrillation. Europace. 2018;20(1):e1–e160. - DOI
-
- Velagić V, de Asmundis C, Mugnai G, Hünük B, Hacioğlu E, Ströker E, et al. Learning curve using the second-generation cryoballoon ablation. J Cardiovasc Med (Hagerstown). 2017;18(7):518–27. - DOI
-
- Bollmann A, Ueberham L, Schuler E, Wiedemann M, Reithmann C, Sause A, et al. Cardiac tamponade in catheter ablation of atrial fibrillation: German-wide analysis of 21 141 procedures in the Helios Atrial Fibrillation Ablation Registry (SAFER). Europace. 2018;20(12):1944–51. - DOI
-
- Mugnai G, de Asmundis C, Hünük B, Ströker E, Moran D, Hacioglu E, et al. Improved visualisation of real-time recordings during third generation cryoballoon ablation: a comparison between the novel short-tip and the second generation device. J Interv Card Electrophysiol. 2016;46(3):307–14. - DOI
-
- Ciconte G, Mugnai G, Sieira J, Velagić V, Saitoh Y, Irfan G, et al. On the quest for the best freeze: predictors of late pulmonary vein reconnections after second-generation cryoballoon ablation. Circ Arrhythm Electrophysiol. 2015;8(6):1359–65. - DOI
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
Research Materials