. 2020 Aug 1;41(29):2731-2742.
doi: 10.1093/eurheartj/ehaa544.

Long-term outcomes after transcatheter aortic valve implantation in failed bioprosthetic valves

Sabine Bleiziffer  1 Matheus Simonato  2 John G Webb  3 Josep Rodés-Cabau  4 Philippe Pibarot  4 Ran Kornowski  5 Stephan Windecker  6 Magdalena Erlebach  7 Alison Duncan  8 Moritz Seiffert  9 Axel Unbehaun  10 Christian Frerker  11 Lars Conzelmann  12 Harindra Wijeysundera  13 Won-Keun Kim  14 Matteo Montorfano  15 Azeem Latib  16 Didier Tchetche  17 Abdelhakim Allali  18 Mohamed Abdel-Wahab  19 Katia Orvin  5 Stefan Stortecky  6 Henrik Nissen  20 Andreas Holzamer  21 Marina Urena  22 Luca Testa  23 Marco Agrifoglio  24 Brian Whisenant  25 Janarthanan Sathananthan  3 Massimo Napodano  26 Antonio Landi  26 Claudia Fiorina  27 Armin Zittermann  1 Verena Veulemans  28 Jan-Malte Sinning  29 Francesco Saia  30 Stephen Brecker  31 Patrizia Presbitero  32 Ole De Backer  33 Lars Søndergaard  33 Giuseppe Bruschi  34 Luis Nombela Franco  35 Anna Sonia Petronio  36 Marco Barbanti  37 Alfredo Cerillo  38 Konstantinos Spargias  39 Joachim Schofer  40 Mauricio Cohen  41 Antonio Muñoz-Garcia  42 Ariel Finkelstein  43 Matti Adam  11 Vicenç Serra  44 Rui Campante Teles  45 Didier Champagnac  46 Alessandro Iadanza  47 Piotr Chodor  48 Holger Eggebrecht  49 Robert Welsh  50 Adriano Caixeta  51 Stefano Salizzoni  52 Antonio Dager  53 Vincent Auffret  54 Asim Cheema  55 Timm Ubben  56 Marco Ancona  15 Tanja Rudolph  1 Jan Gummert  1 Elaine Tseng  57 Stephane Noble  58 Matjaz Bunc  59 David Roberts  60 Malek Kass  61 Anuj Gupta  62 Martin B Leon  63 Danny Dvir  64   65
Affiliations
Free article

Long-term outcomes after transcatheter aortic valve implantation in failed bioprosthetic valves

Sabine Bleiziffer et al. Eur Heart J. .
Free article

Abstract

Aims: Due to bioprosthetic valve degeneration, aortic valve-in-valve (ViV) procedures are increasingly performed. There are no data on long-term outcomes after aortic ViV. Our aim was to perform a large-scale assessment of long-term survival and reintervention after aortic ViV.

Methods and results: A total of 1006 aortic ViV procedures performed more than 5 years ago [mean age 77.7 ± 9.7 years; 58.8% male; median STS-PROM score 7.3% (4.2-12.0)] were included in the analysis. Patients were treated with Medtronic self-expandable valves (CoreValve/Evolut, Medtronic Inc., Minneapolis, MN, USA) (n = 523, 52.0%), Edwards balloon-expandable valves (EBEV, SAPIEN/SAPIEN XT/SAPIEN 3, Edwards Lifesciences, Irvine, CA, USA) (n = 435, 43.2%), and other devices (n = 48, 4.8%). Survival was lower at 8 years in patients with small-failed bioprostheses [internal diameter (ID) ≤ 20 mm] compared with those with large-failed bioprostheses (ID > 20 mm) (33.2% vs. 40.5%, P = 0.01). Independent correlates for mortality included smaller-failed bioprosthetic valves [hazard ratio (HR) 1.07 (95% confidence interval (CI) 1.02-1.13)], age [HR 1.21 (95% CI 1.01-1.45)], and non-transfemoral access [HR 1.43 (95% CI 1.11-1.84)]. There were 40 reinterventions after ViV. Independent correlates for all-cause reintervention included pre-existing severe prosthesis-patient mismatch [subhazard ratio (SHR) 4.34 (95% CI 1.31-14.39)], device malposition [SHR 3.75 (95% CI 1.36-10.35)], EBEV [SHR 3.34 (95% CI 1.26-8.85)], and age [SHR 0.59 (95% CI 0.44-0.78)].

Conclusions: The size of the original failed valve may influence long-term mortality, and the type of the transcatheter valve may influence the need for reintervention after aortic ViV.

Keywords: Aortic valve-in-valve; Reintervention; SAPIEN valve; Severe prosthesis–patient mismatch; TAVR.

PubMed Disclaimer

Comment in

Publication types