A comparative study on the physicochemical characteristics of nanoparticles released in vivo from CoCrMo tapers and cement-stem interfaces of total hip replacements
- PMID: 32596955
- DOI: 10.1002/jbm.b.34667
A comparative study on the physicochemical characteristics of nanoparticles released in vivo from CoCrMo tapers and cement-stem interfaces of total hip replacements
Abstract
The good biocompatibility and corrosion resistance of the bulk CoCrMo alloy has resulted in it being used in the manufacture of implants and load bearing medical devices. These devices, however, can release wear and corrosion products which differ from the composition of the bulk CoCrMo alloy. The physicochemical characteristics of the particles and the associated in vivo reactivity are dictated by the wear mechanisms and electrochemical conditions at the sites of material loss. Debris released from CoCrMo hip bearings, taper junctions, or cement-stem interfaces can, therefore, have different chemical and morphological characteristics, which provide them with different in vivo toxicities. Here, we propose to assess and compare the characteristics of the particles released in vivo from CoCrMo tapers and cement-stem interfaces which have received less attention compared to debris originating from the hip bearings. The study uses state-of-art characterization techniques to provide a detailed understanding of the size, morphology, composition, and chemistry of the particles liberated from the wear and corrosion flakes from revised hip replacements, with an enzymatic treatment. The phase analyses identified Cr2 O3 nanoparticles released from tapers and cement-stem interfaces, whose composition did not vary with origin or particle morphology. The size distributions showed significantly smaller particles were released from the stems, compared to the particles originating from the corresponding tapers. The investigation demonstrates that the tribocorrosive processes occurring at the taper and stem interfaces both result in Cr2 O3 nanoparticle formation.
Keywords: TEM; cobalt chromium; hip replacements; nanoparticles; orthopedics.
© 2020 Wiley Periodicals LLC.
References
REFERENCES
-
- Billi, F., Benya, P., Kavanaugh, A., Adams, J., McKellop, H., & Ebramzadeh, E. (2012). The John Charnley award: An accurate and extremely sensitive method to separate, display, and characterize wear debris. Part 2: Metal and ceramic particles. Clinical Orthopaedics and Related Research, 470(2), 339-350.
-
- Billi, F., Benya, P., Ebramzadeh, E., Campbell, P., Chan, F., & McKellop, H. A. (2009). Metal wear particles: What we know, what we do not know, and why. SAS Journal, 3(4), 133-142. http://dx.doi.org/10.1016/j.esas.2009.11.006.
-
- Bolland, B., Culliford, D. J., Langton, D. J., Millington, J. P., Arden, N. K., & Latham, J. M. (2011). High failure rates with a large-diameter hybrid metal-on-metal total hip replacement: Clinical, radiological and retrieval analysis. The Journal of Bone and Joint Surgery. British Volume, 93(5), 608-615.
-
- Bradberry, S., Wilkinson, J., & Ferner, R. (2014). Systemic toxicity related to metal hip prostheses. Clinical Toxicology, 52(8), 837-847.
-
- Brown, C., Williams, S., Tipper, J. L., Fisher, J., & Ingham, E. (2007). Characterisation of wear particles produced by metal on metal and ceramic on metal hip prostheses under standard and microseparation simulation. Journal of Materials Science: Materials in Medicine, 18(5), 819-827.