Initial multicenter experience with a new high-density coloring module: impact for complex atrial arrhythmias interpretation
- PMID: 32621214
- DOI: 10.1007/s10840-020-00802-1
Initial multicenter experience with a new high-density coloring module: impact for complex atrial arrhythmias interpretation
Abstract
Background: High-density automated mapping of complex atrial tachycardias (ATs) requires accurate assessment of activation maps. A new local activation display module (HD coloring, Biosense Webster®) provides higher map resolution, a better delineation of potential block reducing color interpolation, and a new propagation display. We evaluated the accuracy of a dedicated local activation display compared with standard algorithm.
Methods: High-density maps from 10 AT were collected with a multipolar catheter and were displayed with standard activation or HD coloring. Six expert operators retrospectively analyzed activation maps and were asked to define (1) the tachycardia mechanism, (2) ablation target, and (3) level of difficulty to interpret those maps.
Results: Using HD coloring, operators were able to reach a correct diagnosis in 93% vs. 63%, p < 0.05 compared to standard activation maps. Time to diagnosis was shorter 1.9 ± 1.0 min vs. 3.9 ± 2.1 min, p < 0.05. Confidence level would have allowed ablation without necessity for entrainment maneuvers in 87% vs. 53%, p < 0.05. Operators would have needed to remap or proceed with multiple entrainments in 3% vs. 13% of cases, p < 0.05. Finally, ablation strategy was more accurately identified in 97% vs. 67%, p < 0.05.
Conclusion: Activation mapping with the new HD coloring module allowed a more accurate, reliable, and faster interpretation of complex ATs mechanisms compared to standard activation maps.
Keywords: Ablation; Atrial tachycardia; High-resolution mapping.
References
-
- Durrer D, Roos JP. Epicardial excitation of the ventricles in a patient with wolff-Parkinson-White syndrome (type B). Circulation. 1967;35:15–21.
-
- Kottkamp H, Hindricks G, Breithardt G, Borggrefe M. Three-dimensional electromagnetic catheter technology: electroanatomical mapping of the right atrium and ablation of ectopic atrial tachycardia. J Cardiovasc Electrophysiol. 1997;8:1332–7. - DOI
-
- Shah DC, Jaïs P, Haïssaguerre M, Chouairi S, Takahashi A, Hocini M, et al. Three-dimensional mapping of the common atrial flutter circuit in the right atrium. Circulation. 1997;96:3904–12. - DOI
-
- Takigawa M, Derval N, Frontera A, Martin R, Yamashita S, Cheniti G, et al. Revisiting anatomic macroreentrant tachycardia after atrial fibrillation ablation using ultrahigh-resolution mapping: implications for ablation. Heart Rhythm. 2018;15:326–33. - DOI
-
- Scherr D, Khairy P, Miyazaki S, Aurillac-Lavignolle V, Pascale P, Wilton SB, et al. Five-Year Outcome of Catheter Ablation of Persistent Atrial Fibrillation Using Termination of Atrial Fibrillation as a Procedural Endpoint. Circ-Arrhythmia Elec 205;8:18–24.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources