Implantable cardiac defibrillator leads dysfunction after LVAD implantation
- PMID: 32627211
- DOI: 10.1111/pace.14004
Implantable cardiac defibrillator leads dysfunction after LVAD implantation
Abstract
Background: Implantable cardioverter-defibrillator (ICD) lead dysfunction has been reported after left ventricular assist device (LVAD) implantation in limited single-center studies. We aimed at describing and characterizing the incidence of ICD lead parameters dysfunction after LVAD implantation.
Methods: Among the 652 patients enrolled in the ASSIST-ICD study, only patients with an ICD prior to LVAD were included (n = 401). ICD lead parameters dysfunction following LVAD implantation is defined as follows: (a) >50% decrease in sensing threshold, (b) pacing lead impedance increase/decrease by >100Ω, and (c) >50% increase in pacing threshold.
Results: One hundred twenty-two patients with an ICD prior to LVAD had available ICD interrogation reports prior and after LVAD. A total of 67 (55%) patients exhibited at least one significant lead dysfunction: 17 (15%) exhibited >50% decrease in right ventricular (RV) sensing, 51 (42%) had >100 Ω increase/decrease in RV pacing impedance, and 24 (20%) experienced >50% increase in RV pacing threshold. A total of 52 patients experienced ventricular arrhythmia during follow-up and all were successfully detected and treated by the device. All lead dysfunction could be managed conservatively.
Conclusion: More than 50% of LVAD-recipients may experience >1 significant change in lead parameters but none had severe clinical consequences.
Keywords: implantable cardioverter defibrillator; lead parameters dysfunction; left ventricular assist device.
© 2020 Wiley Periodicals LLC.
References
REFERENCES
-
- Foo D, Walker BD, Kuchar DL, et al. Left ventricular mechanical assist devices and cardiac device interactions: an observational case series. Pacing Clin Electrophysiol. 2009;32:879-887.
-
- Ambardekar AV, Lowery CM, Allen LA, et al. Effect of left ventricular assist device placement on preexisting implantable cardioverter-defibrillator leads. J Card Fail. 2010;16:327-331.
-
- Boudghène-Stambouli F, Boulé S, Goéminne C, et al. Clinical implications of left ventricular assist device implantation in patients with an implantable cardioverter-defibrillator. J Interv Card Electrophysiol. 2014;39:177-184.
-
- Thomas IC, Cork DP, Levy A, et al. ICD lead parameters, performance, and adverse events following continuous-flow LVAD implantation. Pacing Clin Electrophysiol. 2014;37:464-472.
-
- Hu YL, Kasirajan V, Tang DG, et al. Prospective evaluation of implantable cardioverter-defibrillator lead function during and after left ventricular assist device implantation. JACC Clin Electrophysiol. 2016;2:343-354.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical