Non-stationary spatial filtering and accelerated curve fitting for parametric imaging with dynamic PET
- PMID: 3265103
- DOI: 10.1007/BF00252392
Non-stationary spatial filtering and accelerated curve fitting for parametric imaging with dynamic PET
Abstract
A non-stationary spatial low pass filter was developed and implemented in combination with an accelerated non-linear curve fitting routine to create low noise-high contrast images of physiological parameters with dynamic positron emission tomography. The method was applied to 18F-2-fluoro-2-deoxyglucose (FDG) studies, and images of local blood volume, kinetic rate constants, precursor pool volume and glucose metabolism were generated. Noise reduction and contrast preservation was demonstrated in a simulated pie phantom and a study of a patient with a recent brain infarct. Considerably improvement in quantitative accuracy of pixel parameter values was observed in the phantom study in comparison with unprocessed or conventionally smoothed images.
Similar articles
-
Parametric images of myocardial metabolic rate of glucose generated from dynamic cardiac PET and 2-[18F]fluoro-2-deoxy-d-glucose studies.J Nucl Med. 1991 Apr;32(4):733-8. J Nucl Med. 1991. PMID: 2013815
-
Improved positron emission tomography quantification by Fourier-based restoration filtering.Eur J Nucl Med. 1992;19(11):925-32. doi: 10.1007/BF00175857. Eur J Nucl Med. 1992. PMID: 1425779
-
A procedure for wall detection in [18F]FDG positron emission tomography heart studies.Eur J Nucl Med. 1996 Jan;23(1):18-24. doi: 10.1007/BF01736985. Eur J Nucl Med. 1996. PMID: 8586097
-
Does fluorine-18 fluorodeoxyglucose metabolic imaging of tumours benefit oncology?Eur J Nucl Med. 1997 Jun;24(6):691-705. doi: 10.1007/BF00841411. Eur J Nucl Med. 1997. PMID: 9169580 Review.
-
[18F-fluorodeoxyglucose positron emission tomography in Parkinson's disease].Nihon Rinsho. 1997 Jan;55(1):222-6. Nihon Rinsho. 1997. PMID: 9014454 Review. Japanese.
Cited by
-
FDG transport and phosphorylation in human gliomas measured with dynamic PET.J Neurooncol. 1992 Feb;12(2):159-65. doi: 10.1007/BF00172667. J Neurooncol. 1992. PMID: 1560262
-
Nonlinear spatio-temporal filtering of dynamic PET data using a four-dimensional Gaussian filter and expectation-maximization deconvolution.Phys Med Biol. 2013 Feb 21;58(4):1151-68. doi: 10.1088/0031-9155/58/4/1151. Phys Med Biol. 2013. PMID: 23370699 Free PMC article.
-
A gradient-based method for segmenting FDG-PET images: methodology and validation.Eur J Nucl Med Mol Imaging. 2007 Sep;34(9):1427-38. doi: 10.1007/s00259-006-0363-4. Epub 2007 Mar 13. Eur J Nucl Med Mol Imaging. 2007. PMID: 17431616