Free-breathing whole-heart multi-slice myocardial T1 mapping in 2 minutes
- PMID: 32662908
- DOI: 10.1002/mrm.28402
Free-breathing whole-heart multi-slice myocardial T1 mapping in 2 minutes
Abstract
Purpose: To develop and validate a saturation-delay-inversion recovery preparation, slice tracking and multi-slice based sequence for measuring whole-heart native T1 .
Method: The proposed free-breathing sequence performs T1 mapping of multiple left-ventricular slices by slice-interleaved acquisition to collect 10 electrocardiogram-triggered single-shot slice-selective images for each slice. A saturation-delay-inversion recovery pulse is used for T1 preparation. Prospective slice tracking by the diaphragm navigator and retrospective registration are used to reduce through-plane and in-plane motion, respectively. The proposed sequence was validated in both phantom and human subjects (12 healthy subjects and 15 patients who were referred for a clinical cardiac MR exam) and compared with saturation recovery single-shot acquisition (SASHA) and modified Look-Locker inversion recovery (MOLLI).
Results: Phantom T1 measured by the proposed sequence had excellent agreement (R2 = 0.99) with the ground-truth T1 and was insensitive to heart rate. In both healthy subjects and patients, the proposed sequence yielded nine left-ventricular T1 maps per volume in less than 2 minutes (healthy volunteers: 1.8 ± 0.4 minutes; patients: 1.9 ± 0.2 minutes). The average T1 of whole left ventricle for all healthy subjects and patients were 1560 ± 61 and 1535 ± 49 ms by SASHA, 1208 ± 42 and 1233 ± 56 ms by MOLLI5(3)3, and 1397 ± 34 and 1433 ± 56 ms by the proposed sequence, respectively. The corresponding coefficient of variation of T1 were 6.2 ± 1.4% and 5.8 ± 1.6%, 5.3 ± 1.1% and 5.1 ± 0.8%, and 4.9 ± 0.8% and 4.5 ± 0.8%, respectively.
Conclusion: The proposed sequence enables quantification of whole heart T1 with good accuracy and precision in less than 2 minutes during free breathing.
Keywords: free breathing; inversion; myocardial T1 mapping; saturation; whole-heart coverage.
© 2020 International Society for Magnetic Resonance in Medicine.
References
REFERENCES
-
- Messroghli DR, Moon JC, Ferreira VM, et al. Clinical recommendations for cardiovascular magnetic resonance mapping of T1, T2, T2* and extracellular volume: a consensus statement by the Society for Cardiovascular Magnetic Resonance (SCMR) endorsed by the European Association for Cardiovascular Imaging (EACVI). J Cardiovasc Magn Reson. 2017;19:75.
-
- Moon JC, Messroghli DR, Kellman P, et al. Myocardial T1 mapping and extracellular volume quantification: a Society for Cardiovascular Magnetic Resonance (SCMR) and CMR Working Group of the European Society of Cardiology consensus statement. J Cardiovasc Magn Reson. 2013;15:92.
-
- Mewton N, Liu CY, Croisille P, Bluemke D, Lima JA. Assessment of myocardial fibrosis with cardiovascular magnetic resonance. J Am Coll Cardiol. 2011;57:891-903.
-
- Ugander M, Bagi PS, Oki AJ, et al. Myocardial edema as detected by pre-contrast T1 and T2 CMR delineates area at risk associated with acute myocardial infarction. JACC Cardiovasc Imaging. 2012;5:596-603.
-
- Karamitsos TD, Piechnik SK, Banypersad SM, et al. Noncontrast T1 mapping for the diagnosis of cardiac amyloidosis. JACC Cardiovasc Imaging. 2013;6:488-497.