Variplex™ test system fails to reliably detect SARS-CoV-2 directly from respiratory samples without RNA extraction
- PMID: 32681309
- PMCID: PMC7367510
- DOI: 10.1007/s10096-020-03983-9
Variplex™ test system fails to reliably detect SARS-CoV-2 directly from respiratory samples without RNA extraction
Abstract
Diagnosis of COVID is performed by PCR methods, but their capacity is limited by the requirement of high-level facilities and instruments. The loop-mediated isothermal amplification (LAMP) method has been utilized for the detection of isolated virus-specific RNA. Preliminary data suggest the possibility of isothermal amplification directly from respiratory samples without RNA extraction. All patients admitted to our hospital were screened for SARS-CoV-2 by routine. Respiratory samples were tested by variplex system based on LAMP method directly without RNA extraction and by PCR. Primary endpoint was the false-negative rate of variplex test compared with PCR as gold standard. In 109 patients variplex test and PCR assay were performed simultaneously. Median age was 80 years and male/female ratio was 40/60%. The prevalence of PCR-confirmed COVID diagnosis was 43.1%. Variplex test was positive in 13.8%. False-negative rate of variplex test compared with PCR was 83.0%. The potential of LAMP technology using isolated RNA has been demonstrated impressively by others, and excellent sensitivity and specificity of detecting SARS-CoV-2 has been reported. However, without RNA extraction, the variplex test system failed to reliably detect SARS-CoV-2 directly in respiratory samples.
Keywords: COVID-19 diagnostic testing; False-negative reactions; Infectious disease outbreaks; Loop-mediated isothermal amplification; SARS-CoV-2; Specificity and sensitivity.
Conflict of interest statement
The authors declare that they have no conflict of interest.
Similar articles
-
Development and Clinical Application of a Rapid and Sensitive Loop-Mediated Isothermal Amplification Test for SARS-CoV-2 Infection.mSphere. 2020 Aug 26;5(4):e00808-20. doi: 10.1128/mSphere.00808-20. mSphere. 2020. PMID: 32848011 Free PMC article.
-
Validation of a single-step, single-tube reverse transcription loop-mediated isothermal amplification assay for rapid detection of SARS-CoV-2 RNA.J Med Microbiol. 2020 Sep;69(9):1169-1178. doi: 10.1099/jmm.0.001238. Epub 2020 Jul 31. J Med Microbiol. 2020. PMID: 32755529 Free PMC article.
-
Loop mediated isothermal amplification (LAMP) assays as a rapid diagnostic for COVID-19.Med Hypotheses. 2020 Aug;141:109786. doi: 10.1016/j.mehy.2020.109786. Epub 2020 Apr 25. Med Hypotheses. 2020. PMID: 32361529 Free PMC article.
-
Detection of COVID-19: A review of the current literature and future perspectives.Biosens Bioelectron. 2020 Oct 15;166:112455. doi: 10.1016/j.bios.2020.112455. Epub 2020 Jul 21. Biosens Bioelectron. 2020. PMID: 32739797 Free PMC article. Review.
-
Coronavirus disease 2019: What we know?J Med Virol. 2020 Jul;92(7):719-725. doi: 10.1002/jmv.25766. Epub 2020 Mar 28. J Med Virol. 2020. PMID: 32170865 Free PMC article. Review.
Cited by
-
A lab-on-a-chip platform for integrated extraction and detection of SARS-CoV-2 RNA in resource-limited settings.Anal Chim Acta. 2021 Sep 8;1177:338758. doi: 10.1016/j.aca.2021.338758. Epub 2021 Jun 14. Anal Chim Acta. 2021. PMID: 34482896 Free PMC article.
-
SARS-CoV-2 Diagnostics Based on Nucleic Acids Amplification: From Fundamental Concepts to Applications and Beyond.Front Cell Infect Microbiol. 2022 Mar 23;12:799678. doi: 10.3389/fcimb.2022.799678. eCollection 2022. Front Cell Infect Microbiol. 2022. PMID: 35402302 Free PMC article. Review.
-
Laboratory-based molecular test alternatives to RT-PCR for the diagnosis of SARS-CoV-2 infection.Cochrane Database Syst Rev. 2024 Oct 14;10(10):CD015618. doi: 10.1002/14651858.CD015618. Cochrane Database Syst Rev. 2024. PMID: 39400904
-
Tools and Techniques for Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2)/COVID-19 Detection.Clin Microbiol Rev. 2021 May 12;34(3):e00228-20. doi: 10.1128/CMR.00228-20. Print 2021 Jun 16. Clin Microbiol Rev. 2021. PMID: 33980687 Free PMC article. Review.
-
Advances in laboratory detection methods and technology application of SARS-CoV-2.J Med Virol. 2022 Apr;94(4):1357-1365. doi: 10.1002/jmv.27494. Epub 2021 Dec 10. J Med Virol. 2022. PMID: 34854101 Free PMC article. Review.
References
-
- Zhu N, Zhang D, Wang W, Li X, Yang B, Song J, Zhao X, Huang B, Shi W, Lu R, Niu P, Zhan F, Ma X, Wang D, Xu W, Wu G, Gao GF, Tan W, China Novel Coronavirus Investigating and Research Team A novel coronavirus from patients with pneumonia in China, 2019. N Engl J Med. 2020;382:727–733. doi: 10.1056/NEJMoa2001017. - DOI - PMC - PubMed
-
- Wu F, Zhao S, Yu B, Chen Y-M, Wang W, Song Z-G, Hu Y, Tao Z-W, Tian J-H, Pei Y-Y, Yuan M-L, Zhang Y-L, Dai F-H, Liu Y, Wang Q-M, Zheng J-J, Xu L, Holmes EC, Zhang Y-Z. A new coronavirus associated with human respiratory disease in China. Nature. 2020;579:265–269. doi: 10.1038/s41586-020-2008-3. - DOI - PMC - PubMed
-
- Huang C, Wang Y, Li X, Ren L, Zhao J, Hu Y, Zhang L, Fan G, Xu J, Gu X, Cheng Z, Yu T, Xia J, Wei Y, Wu W, Xie X, Yin W, Li H, Liu M, Xiao Y, Gao H, Guo L, Xie J, Wang G, Jiang R, Gao Z, Jin Q, Wang J, Cao B. Clinical features of patients infected with 2019 novel coronavirus in Wuhan, China. Lancet Lond Engl. 2020;395:497–506. doi: 10.1016/S0140-6736(20)30183-5. - DOI - PMC - PubMed
MeSH terms
Substances
Supplementary concepts
LinkOut - more resources
Full Text Sources
Medical
Miscellaneous