. 2020 Dec:80:105104.
doi: 10.1016/j.clinbiomech.2020.105104. Epub 2020 Jul 16.

Biomechanical in vitro examination of a standardized low-volume tubular femoroplasty

Affiliations
Free article

Biomechanical in vitro examination of a standardized low-volume tubular femoroplasty

Andreas J Horbach et al. Clin Biomech (Bristol). 2020 Dec.
Free article

Abstract

Background: Osteoporosis is associated with the risk of fractures near the hip. Age and comorbidities increase the perioperative risk. Due to the ageing population, fracture of the proximal femur also proves to be a socio-economic problem. Preventive surgical measures have hardly been used so far.

Methods: 10 pairs of human femora from fresh cadavers were divided into control and low-volume femoroplasty groups and subjected to a Hayes fall-loading fracture test. The results of the respective localization and classification of the fracture site, the Singh index determined by computed tomography (CT) examination and the parameters in terms of fracture force, work to fracture and stiffness were evaluated statistically and with the finite element method. In addition, a finite element parametric study with different position angles and variants of the tubular geometry of the femoroplasty was performed.

Findings: Compared to the control group, the work to fracture could be increased by 33.2%. The fracture force increased by 19.9%. The used technique and instrumentation proved to be standardized and reproducible with an average poly(methyl methacrylate) volume of 10.5 ml. The parametric study showed the best results for the selected angle and geometry.

Interpretation: The cadaver studies demonstrated the biomechanical efficacy of the low-volume tubular femoroplasty. The numerical calculations confirmed the optimal choice of positioning as well as the inner and outer diameter of the tube in this setting. The standardized minimally invasive technique with the instruments developed for it could be used in further comparative studies to confirm the measured biomechanical results.

Keywords: Biomechanical test; Femoroplasty; Osteoporosis; Prevention; Proximal femur fracture.