Visualization of flow diverter stent wall apposition during intracranial aneurysm treatment using a virtually diluted cone beam CT technique (Vessel ASSIST)
- PMID: 32740710
- DOI: 10.1007/s00234-020-02507-8
Visualization of flow diverter stent wall apposition during intracranial aneurysm treatment using a virtually diluted cone beam CT technique (Vessel ASSIST)
Abstract
Purpose: Flow diverters (FD) have poor radiopacity, challenging visualization of deployment and vessel wall apposition with conventional neuroimaging modalities. We evaluated a novel cone beam computed tomography (CT) imaging technique that allows virtual dilution (VD) of contrast media to facilitate workflow and ensure accurate assessment of FD wall apposition.
Methods: We retrospectively evaluated all patients treated for intracranial aneurysms with FD at our institution between November 2018 and November 2019. Undiluted injected dual cone beam CT acquisitions performed post-stenting were displayed with VD software (GE Healthcare). The resulting images were compared with conventional two-dimensional (2D) digital subtraction angiography (DSA) images. Two neurointerventionalists (Reader 1 and Reader 2, (R1, R2)) independently assessed FD deployment and wall apposition. Confidence in the diagnosis, inter-reader agreement, and X-ray exposure were assessed.
Results: A total of 27 cases were reviewed. FD deployment and wall apposition scores were 4.2 ± 1.0 (R1) and 4.0 ± 1.1 (R2) for DSA and 3.7 ± 1.2 (R1) and 4.1 ± 1.0 (R2) for VD. Confidence in the diagnosis was improved with VD, with scores of 3.7 ± 0.7 (R1) and 4.0 ± 0.7 (R2) using DSA and 4.9 ± 0.2 (R1) and 4.9 ± 0.2 (R2) using VD (P < 0.001). Inter-reader agreement using 2D DSA was improved from moderate (0.49324) to good (0.7272) (P < 0.0001). There were no significant differences in inter-reader agreement in the deployment assessment (P = 0.68) or dose-area product (P = 0.54) between techniques.
Conclusion: VD imaging with dual cone beam CT enables accurate assessment of FD wall apposition after deployment with greater confidence and improved inter-reader agreement versus conventional 2D DSA alone, with comparable X-ray exposure.
Keywords: Cone beam CT; Flow diverter; Imaging; Intracranial aneurysm; Virtual dilution.
Similar articles
-
Clinical evaluation of high-resolution cone-beam computed tomography for the implantation of flow-diverter stents in intracranial aneurysms.J Clin Neurosci. 2022 Sep;103:14-19. doi: 10.1016/j.jocn.2022.06.028. Epub 2022 Jul 5. J Clin Neurosci. 2022. PMID: 35797769
-
Flow Diverter Apposition in Patients with Large or Giant Intracranial Aneurysms Evaluated on Three-Dimensional Fusion Images Acquired by High-Resolution Cone-Beam Computed Tomography and Digital Subtraction Angiography.World Neurosurg. 2021 Mar;147:e388-e395. doi: 10.1016/j.wneu.2020.12.068. Epub 2021 Jan 7. World Neurosurg. 2021. PMID: 33359518
-
A technical note on intra-arterial cone-beam computed tomography for the evaluation of flow-diverter stents: Image quality differences between diluted and non-diluted contrast medium.Interv Neuroradiol. 2020 Apr;26(2):164-169. doi: 10.1177/1591019919890929. Epub 2019 Nov 25. Interv Neuroradiol. 2020. PMID: 31766961 Free PMC article.
-
Use of self-expanding stents for better intracranial flow diverter wall apposition.Interv Neuroradiol. 2017 Apr;23(2):129-136. doi: 10.1177/1591019916681981. Epub 2016 Dec 12. Interv Neuroradiol. 2017. PMID: 27956518 Free PMC article.
-
Flat detector CT and its applications in the endovascular treatment of wide-necked intracranial aneurysms-A literature review.Eur J Radiol. 2017 Mar;88:26-31. doi: 10.1016/j.ejrad.2016.12.027. Epub 2016 Dec 26. Eur J Radiol. 2017. PMID: 28189205 Review.
Cited by
-
Cone-beam CT angiography to assess the microvascular anatomy of intracranial arterial dissections.Neuroradiol J. 2022 Aug;35(4):527-532. doi: 10.1177/19714009211049089. Epub 2021 Sep 30. Neuroradiol J. 2022. PMID: 34590889 Free PMC article.
-
A Comparability Study Between Intravenous Contrast-Enhanced Cone-Beam Computed Tomography (CBCT) and Magnetic Resonance Angiography (MRA) on the Post-Treatment Follow-Up of Intracranial Aneurysms: A Single-Center Prospective Cohort Study.Diagnostics (Basel). 2025 Jul 14;15(14):1774. doi: 10.3390/diagnostics15141774. Diagnostics (Basel). 2025. PMID: 40722523 Free PMC article.
References
-
- Kallmes DF, Hanel R, Lopes D, Boccardi E, Bonafe A, Cekirge S, Fiorella D, Jabbour P, Levy E, McDougall C, Siddiqui A, Szikora I, Woo H, Albuquerque F, Bozorgchami H, Dashti SR, Delgado Almandoz JE, Kelly ME, Turner R, Woodward BK, Brinjikji W, Lanzino G, Lylyk P (2015) International retrospective study of the pipeline embolization device: a multicenter aneurysm treatment study. AJNR Am J Neuroradiol 36(1):108–115. https://doi.org/10.3174/ajnr.A4111 - DOI - PubMed
-
- Brinjikji W, Murad MH, Lanzino G, Cloft HJ, Kallmes DF (2013) Endovascular treatment of intracranial aneurysms with flow diverters: a meta-analysis. Stroke 44(2):442–447. https://doi.org/10.1161/STROKEAHA.112.678151 - DOI - PubMed
-
- Kizilkilic O, Kocer N, Metaxas GE, Babic D, Homan R, Islak C (2012) Utility of VasoCT in the treatment of intracranial aneurysm with flow-diverter stents. J Neurosurg 117(1):45–49. https://doi.org/10.3171/2012.4.JNS111660 - DOI - PubMed
-
- Ding D, Starke RM, Durst CR, Gaughen JR Jr, Evans AJ, Jensen ME, Liu KC (2014) DynaCT imaging for intraprocedural evaluation of flow-diverting stent apposition during endovascular treatment of intracranial aneurysms. J Clin Neurosci 21(11):1981–1983. https://doi.org/10.1016/j.jocn.2014.04.003 - DOI - PubMed
-
- Heller R, Calnan DR, Lanfranchi M, Madan N, Malek AM (2013) Incomplete stent apposition in Enterprise stent-mediated coiling of aneurysms: persistence over time and risk of delayed ischemic events. J Neurosurg 118(5):1014–1022. https://doi.org/10.3171/2013.2.JNS121427 - DOI - PubMed
MeSH terms
LinkOut - more resources
Full Text Sources
Medical