Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2021 Apr;225(2):155-160.
doi: 10.1055/a-1209-3861. Epub 2020 Aug 3.

Experience with Real-Time Continuous Glucose Monitoring in Newborns with Congenital Hyperinsulinemic Hypoglycemia

Affiliations

Experience with Real-Time Continuous Glucose Monitoring in Newborns with Congenital Hyperinsulinemic Hypoglycemia

Ayşe Anık et al. Z Geburtshilfe Neonatol. 2021 Apr.

Abstract

Background: Effective treatment and close monitoring of hypoglycemia in children with congenital hyperinsulinemic hypoglycemia (CHH) is vital to prevent brain damage. The current use of capillary sampling schedules does not provide a comprehensive assessment of glycemic status and fails to detect asymptomatic hypoglycemia episodes.

Aim: To investigate the efficacy and accuracy of a real-time continuous glucose monitoring system (RT-CGMS) in neonates with CHH.

Methods: A sensor connected to RT-CGMS was inserted into the newborn patients and maintained for at least 6 days during their stay in the hospital. We compared the readings of CGMS with capillary blood glucose values using Bland-Altman analysis.

Results: A total of 110 blood glucose values were compared to readings from the CGMS. All results were calculated and plotted for CGMS values at 0-4, 5-9, 10-14, 15-19, 20-24, and 25-29 min after capillary blood glucose sampling. CGMS readings were highly correlated with blood glucose values, especially during normoglycemia. In case of hypoglycemia, the mean difference between the CGMS and capillary glucose values was higher. Although the false positive rate for hypoglycemia was relatively high in CGMS, RT-CGMS may show some episodes of hypoglycemia earlier than blood measurement.

Conclusion: RT-CGMS is accurate during normoglycemia, and it can reduce the number of capillary blood samples in children with CHH.