Evaluation of the quality of bedside monitoring of the blood glucose level in a teaching hospital
- PMID: 3275478
- PMCID: PMC1267502
Evaluation of the quality of bedside monitoring of the blood glucose level in a teaching hospital
Abstract
We evaluated the precision and accuracy of monitoring of the blood glucose level in the laboratory and at the bedside with one of four glucose meters by an experienced operator and by 39 nurses in a teaching hospital. For precision studies aqueous quality control materials were used. A total of 85 blood samples were tested. The precision of the glucose meters (expressed as the coefficient of variation [CV]) in the hands of the experienced operator ranged from 6.7% to 11.1%. The correlation between the values obtained by the experienced operator and the reference values obtained in the laboratory was high (0.95 to 0.98). The precision of the values obtained by the nurses using the meters ranged from 13.7% on medical wards to 45.7% in the intensive care unit (ICU). The correlation between these values and those obtained in the central laboratory ranged from 0.72 to 0.82. Twenty-four percent of the glucose values determined on medical wards and 62% of those determined in the ICU deviated from the reference value by at least 20%. Of the 85 patients 12 (14%) would have received different insulin dosages had the reference value been available at the same time as the glucose meter reading: in 3 of the patients the discrepancy was 6 units of insulin or greater. Continuous quality control of bedside monitoring of the blood glucose level is needed. In addition, personnel who use glucose meters should receive adequate training.
Similar articles
-
A hospital-wide blind control program for bedside glucose meters.Arch Pathol Lab Med. 1989 Dec;113(12):1370-5. Arch Pathol Lab Med. 1989. PMID: 2589948
-
Accuracy and feasibility of point-of-care and continuous blood glucose analysis in critically ill ICU patients.Crit Care. 2006;10(5):R135. doi: 10.1186/cc5048. Crit Care. 2006. PMID: 16981981 Free PMC article.
-
Interinstitutional comparison of bedside blood glucose monitoring program characteristics, accuracy performance, and quality control documentation: a College of American Pathologists Q-Probes study of bedside blood glucose monitoring performed in 226 small hospitals.Arch Pathol Lab Med. 1998 Jun;122(6):495-502. Arch Pathol Lab Med. 1998. PMID: 9625416
-
Glucose control in the intensive care unit.Crit Care Med. 2009 May;37(5):1769-76. doi: 10.1097/CCM.0b013e3181a19ceb. Crit Care Med. 2009. PMID: 19325461 Review.
-
Glucose meter accuracy and the impact on the care of diabetes in childhood and adolescence.Pediatr Endocrinol Rev. 2011 Mar;8(3):200-7. Pediatr Endocrinol Rev. 2011. PMID: 21525796 Review.
Cited by
-
The need for continuous blood glucose monitoring in the intensive care unit.J Diabetes Sci Technol. 2007 May;1(3):412-4. doi: 10.1177/193229680700100315. J Diabetes Sci Technol. 2007. PMID: 19885098 Free PMC article.
-
Bedside monitoring of the blood glucose level: the high cost of reliable results.CMAJ. 1989 Apr 15;140(8):899-901. CMAJ. 1989. PMID: 2702527 Free PMC article. No abstract available.
-
Evaluation of bedside blood glucose monitoring by the phlebotomy team in an acute general care hospital.Acta Diabetol. 1995 Dec;32(4):213-6. doi: 10.1007/BF00576252. Acta Diabetol. 1995. PMID: 8750758
-
Self monitoring of blood glucose concentrations by non-insulin dependent diabetic patients.BMJ. 1992 Nov 14;305(6863):1171-2. doi: 10.1136/bmj.305.6863.1171. BMJ. 1992. PMID: 1467715 Free PMC article. No abstract available.
References
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical