Design of an automatic spectrophotometric system
- PMID: 32797917
- DOI: 10.1016/j.talanta.2020.121163
Design of an automatic spectrophotometric system
Abstract
A Multi-Syringe Flow Injection Analysis (MSFIA) fluorometric system based on a 3D printing device hosting a CCD detector has been designed for the determination of quinine in soft drinks. A LED controlled by an electronic circuit was used as a radiation source. The entire system is connected to two USB outputs of a computer. The AutoAnalysis program has been used for data acquisition and treatment. The results are more accurate and precise than those obtained with a manual method using a conventional spectrofluorometer. This developed system is a viable alternative to reduce the consumption of reagents, the impact on the environment and reduce measurement costs.
Keywords: 3D printed; CCD detector; Fluorometric system; MSFIA; Quinine; Soft drinks.
Copyright © 2020 Elsevier B.V. All rights reserved.