Comparative Study
. 2020 Sep 14;13(17):2012-2020.
doi: 10.1016/j.jcin.2020.04.054. Epub 2020 Aug 12.

Management and Outcomes of Transvenous Pacing Leads in Patients Undergoing Transcatheter Tricuspid Valve Replacement

Affiliations
Free article
Comparative Study

Management and Outcomes of Transvenous Pacing Leads in Patients Undergoing Transcatheter Tricuspid Valve Replacement

Jason H Anderson et al. JACC Cardiovasc Interv. .
Free article

Abstract

Objectives: The aim of this study was to determine the prevalence of pacemaker lead-related complications following transcatheter tricuspid valve replacement (TTVR).

Background: The rate of permanent pacemaker implantation following tricuspid valve (TV) surgery is high, and many patients have transvenous leads. The feasibility, safety, and outcomes of subsequently performing TTVR in the setting of transvenous pacemaker leads have not been established.

Methods: The VIVID (Valve-in-Valve International Database) registry was used to review 329 patients who underwent TTVR following TV repair or replacement. Patients were subdivided into 3 cohorts for intergroup comparisons: no lead, epicardial lead, and transvenous lead (entrapped or not entrapped during the TTVR procedure).

Results: Of 329 patients who underwent TTVR, 128 (39%) had prior pacing systems in place, 70 with epicardial and 58 with transvenous leads. A total of 31 patients had leads passing through the TV. Three patients had the right ventricular (RV) lead extracted prior to TTVR. The remaining 28 patients had the RV lead entrapped between the transcatheter TV implant and the surgical valve (n = 22) or the repaired TV (n = 6). One patient had displacement of the RV lead during the procedure, and 2 experienced lead failure during follow-up. Overall, there was no significant difference in the cumulative incidences of competing outcomes (death, TV reintervention, TV dysfunction) between patients with and those without pacing leads or entrapped RV leads.

Conclusions: TTVR in the setting of trans-TV pacemaker leads without lead extraction or re-replacement can be performed safely with a low risk for complications, offering an alternative to surgical TV replacement.

Keywords: artificial; heart valves; pacemaker; tricuspid valve.