. 2021 Jul;37(3):427-435.
doi: 10.1007/s11282-020-00474-7. Epub 2020 Aug 16.

Low-dose cone-beam computed tomography in simulated condylar erosion detection: a diagnostic accuracy study

Affiliations

Low-dose cone-beam computed tomography in simulated condylar erosion detection: a diagnostic accuracy study

Noha Saleh Abu-Taleb et al. Oral Radiol. 2021 Jul.

Abstract

Objectives: The purpose of the present study was to assess the diagnostic accuracy of low-dose cone-beam computed tomography (CBCT) in the detection of simulated mandibular condyle erosions.

Methods: 102 simulated erosions were performed on the condyles of eight dry human mandibles. Each mandible was subjected to four CBCT scan protocols: high-definition (HD), normal definition (NORM), ultra-low-dose high-definition (ULD-HD), and ultra-low-dose normal definition (ULD-NORM). All scans were analyzed by two observers. The inter-observer and intra-observer agreement as well as the agreement with the gold standard were assessed. The sensitivity, specificity, positive-predictive value, negative-predictive value and accuracy of erosion detection were calculated.

Results: A substantial to almost perfect agreement with the gold standard was found regarding the HD protocol and substantial agreement in NORM and ULD-HD protocols; however, moderate agreement was found regarding the ULD-NORM protocol. The sensitivity, specificity and accuracy values were highest for the HD protocol followed by the NORM and ULD-HD which showed comparable results; while, the ULD-NORM protocol showed the least values.

Conclusions: The studied ULD-HD CBCT protocol can be recommended for the detection of mandibular condylar erosions due to the reduced radiation dose; however, ULD-NORM is not advocated for similar clinical use.

Keywords: Cone-beam computed tomography; Mandibular condyle; Observer variation; Radiation dosage; Sensitivity and specificity.

PubMed Disclaimer

References

    1. Naser AZ, Shirani AM, Hekmatian E, Valiani A, Ardestani P, Vali A. Comparison of accuracy of uncorrected and corrected sagittal tomography in detection of mandibular condyle erosions. Dent Res J (Isfahan). 2010;7(2):76–81.
    1. Santos LAN, Campos PSF, Paula AMP, Martelli Júnior H, Melo Filho MR. Image of juvenile idiopathic arthritis in mandibular condyle–case report. Rev ABRO. 2005;6(1):29–34.
    1. Tanaka E, Detamore MS, Mercuri LG. Degenerative disorders of the temporomandibular joint: etiology, diagnosis, and treatment. J Dent Res. 2008;87(4):296–307. - DOI
    1. Robinson B, Kelma V, Marques LS, Pereira LJ. Imaging diagnosis of the temporomandibular joint. Oral Radiol. 2009;25(2):86–98. - DOI
    1. Marques AP, Perrella A, Arita ES, Pereira MF, Cavalcanti M. Assessment of simulated mandibular condyle bone lesions by cone beam computed tomography. Braz Oral Res. 2010;24(4):467–74. - DOI