Fast and accurate online calibration of optical see-through head-mounted display for AR-based surgical navigation using Microsoft HoloLens
- PMID: 32809184
- DOI: 10.1007/s11548-020-02246-4
Fast and accurate online calibration of optical see-through head-mounted display for AR-based surgical navigation using Microsoft HoloLens
Abstract
Purpose: The use of an optical see-through head-mounted display (OST-HMD) in augmented reality (AR) has significantly increased in recent years, but the alignment between the virtual scene and physical reality is still a challenge. A fast and accurate calibration method of OST-HMD is important for augmented reality in the medical field.
Methods: We proposed a fast online calibration procedure for OST-HMD with the aid of an optical tracking system. Two 3D datasets are collected in this procedure: the virtual points rendered in front of the observer's eyes and the corresponding points in optical tracking space. The transformation between these two 3D coordinates is solved to build the connection between virtual and real space. An AR-based surgical navigation system is developed with the help of this procedure, which is used for experiment verification and clinical trial.
Results: Phantom experiment based on the 3D-printed skull is performed, and the average root-mean-square error of control points between rendered object and skull model is [Formula: see text] mm, and the time consumption of the calibration procedure can be less than 30 s. A clinical trial is also conducted to show the feasibility in real surgery theatre.
Conclusions: The proposed calibration method does not rely on the camera of the OST-HMD and is very easy to operate. Phantom experiment and clinical case demonstrated the feasibility of our AR-based surgical navigation system and indicated it has the potential in clinical application.
Keywords: Augmented reality; Calibration; Microsoft HoloLens; Surgical navigation.
References
-
- Arun KS, Huang TS, Blostein SD (1987) Least-squares fitting of two 3-d point sets. IEEE Trans Pattern Anal Mach Intell PAMI 9(5):698–700. https://doi.org/10.1109/TPAMI.1987.4767965 - DOI
-
- Azimi E, Qian L, Navab N, Kazanzides P (2017) Alignment of the virtual scene to the tracking space of a mixed reality head-mounted display. arXiv preprint arXiv:1703.05834
-
- Besl PJ, McKay ND (1992) A method for registration of 3-d shapes. IEEE Trans Pattern Anal Mach Intell 14(2):239–256. https://doi.org/10.1109/34.121791 - DOI
-
- Borgmann H, Socarrás MR, Salem J, Tsaur I, Rivas JG, Barret E, Tortolero L (2017) Feasibility and safety of augmented reality-assisted urological surgery using smartglass. World J Urol 35(6):967–972 - DOI
-
- Brown J (2019) Op11 wearable technology and simulation to support learning. is it the way forward? A pilot study in primary care with healthcare professionals across the north west: Wearable technology. BMJ Simul Technol Enhanced Learn 5:5–A7
MeSH terms
Grants and funding
- 2017YFB13029/Key Technology Research and Development Program of Shandong (CN)
- 2017YFB1104100/the National Key R&D Program of China
- 81971709/National Natural Science Foundation of China
- 81828003/National Natural Science Foundation of China
- M-0019/National Natural Science Foundation of China
- 82011530141/National Natural Science Foundation of China
- 2018C01038/the Foundation of Ministry of Education of China Science and Technology Development Center
- 19510712200/the Foundation of Science and Technology Commission of Shanghai Municipality
- ZH20182DA15/Shanghai Jiao Tong University Foundation on Medical and Technological Joint Science Research
- YG2019ZDA06/Shanghai Jiao Tong University Foundation on Medical and Technological Joint Science Research
- ZH2018QNA23/Shanghai Jiao Tong University Foundation on Medical and Technological Joint Science Research
LinkOut - more resources
Full Text Sources
Other Literature Sources
Research Materials