Comparative Study
. 2021 Jan;41(1):114-119.
doi: 10.3343/alm.2021.41.1.114. Epub 2020 Aug 25.

Comparison of Three Blood Collection Tubes for 35 Biochemical Analytes: The Becton Dickinson Barricor Tube, Serum Separating Tube, and Plasma Separating Tube

Affiliations
Comparative Study

Comparison of Three Blood Collection Tubes for 35 Biochemical Analytes: The Becton Dickinson Barricor Tube, Serum Separating Tube, and Plasma Separating Tube

Sunghwan Shin et al. Ann Lab Med. 2021 Jan.

Abstract

The Barricor tube (Becton Dickinson [BD], Sunnyvale, CA, USA) was recently developed to mechanically separate plasma by increasing the centrifugation rate. We compared the Barricor tube with existing serum- and plasma-based tubes based on 35 biochemical analytes and preanalytical turnaround time (TAT). Blood samples were collected from 30 healthy volunteers in a Barricor tube, serum separating tube (SST, Vacutainer SST II Tube 8.5 mL, #368972; BD), or plasma separating tube (PST, Vacutainer PST Tube 8.0 mL, #367964; BD) in random order. Next, 27 chemistry analytes, six immunochemistry analytes, and two cardiac markers were compared using Passing-Bablok regression and the Bland-Altman method. Preanalytical TAT was measured for each tube. The Barricor tube exhibited bias exceeding the desirable limit for nine and four analytes compared with the SST and PST, respectively. The Barricor tube lactate dehydrogenase value showed a bias of -10.29% and -9.86% compared with that of the SST and PST, respectively. The preanalytical TAT of Barricor tube was 8.8 minutes, which was the shortest among the three tubes. The clinical performance of the Barricor tube was equivalent to that of the SST and PST for most analytes, with an apparent advantage in preanalytical TAT. When using the Barricor tube, the reference range needs to be changed for some analytes that exceed the desirable bias limit.

Keywords: Analytes; Barricor tube; Bias; Plasma separating tube; Serum separating tube; Turnaround time.

PubMed Disclaimer

Conflict of interest statement

CONFLICTS OF INTEREST

None declared.

Similar articles

Cited by

References

    1. Fournier JE, Northrup V, Clark C, Fraser J, Howlett M, Atkinson P, et al. Evaluation of BD Vacutainer Barricor blood collection tubes for routine chemistry testing on a Roche Cobas 8000 Platform. Clin Biochem. 2018;58:94–9. - PubMed
    1. Füzéry AK, Raizman JE, Goudreau BL, Moses K, Niemann K, Park J, et al. The BD Barricor blood collection tube is an acceptable and robust alternative to the PST for use with the Beckman AccuTnI+3 assay. Clin Biochem. 2017;50:851–7. - PubMed
    1. Dupuy AM, Badiou S, Daubin D, Bargnoux AS, Magnan C, Klouche K, et al. Comparison of Barricor vs. lithium heparin tubes for selected routine biochemical analytes and evaluation of post centrifugation stability. Biochem Med (Zagreb) 2018;28:020902. - PMC - PubMed
    1. Cadamuro J, Mrazek C, Leichtle AB, Kipman U, Felder TK, Wiedemann H, et al. Influence of centrifugation conditions on the results of 77 routine clinical chemistry analytes using standard vacuum blood collection tubes and the new BD-Barricor tubes. Biochem Med (Zagreb) 2018;28:010704. - PMC - PubMed
    1. Arslan FD, Karakoyun I, Basok BI, Aksit MZ, Baysoy A, Ozturk YK, et al. The local clinical validation of a new lithium heparin tube with a barrier: BD Vacutainer Barricor LH Plasma tube. Biochem Med (Zagreb) 2017;27:030706. - PMC - PubMed

Publication types

MeSH terms

LinkOut - more resources