Comparison of Three Blood Collection Tubes for 35 Biochemical Analytes: The Becton Dickinson Barricor Tube, Serum Separating Tube, and Plasma Separating Tube
- PMID: 32829587
- PMCID: PMC7443522
- DOI: 10.3343/alm.2021.41.1.114
Comparison of Three Blood Collection Tubes for 35 Biochemical Analytes: The Becton Dickinson Barricor Tube, Serum Separating Tube, and Plasma Separating Tube
Abstract
The Barricor tube (Becton Dickinson [BD], Sunnyvale, CA, USA) was recently developed to mechanically separate plasma by increasing the centrifugation rate. We compared the Barricor tube with existing serum- and plasma-based tubes based on 35 biochemical analytes and preanalytical turnaround time (TAT). Blood samples were collected from 30 healthy volunteers in a Barricor tube, serum separating tube (SST, Vacutainer SST II Tube 8.5 mL, #368972; BD), or plasma separating tube (PST, Vacutainer PST Tube 8.0 mL, #367964; BD) in random order. Next, 27 chemistry analytes, six immunochemistry analytes, and two cardiac markers were compared using Passing-Bablok regression and the Bland-Altman method. Preanalytical TAT was measured for each tube. The Barricor tube exhibited bias exceeding the desirable limit for nine and four analytes compared with the SST and PST, respectively. The Barricor tube lactate dehydrogenase value showed a bias of -10.29% and -9.86% compared with that of the SST and PST, respectively. The preanalytical TAT of Barricor tube was 8.8 minutes, which was the shortest among the three tubes. The clinical performance of the Barricor tube was equivalent to that of the SST and PST for most analytes, with an apparent advantage in preanalytical TAT. When using the Barricor tube, the reference range needs to be changed for some analytes that exceed the desirable bias limit.
Keywords: Analytes; Barricor tube; Bias; Plasma separating tube; Serum separating tube; Turnaround time.
Conflict of interest statement
None declared.
Similar articles
-
Barricor blood collection tubes are equivalent to PST for a variety of chemistry and immunoassay analytes except for lactate dehydrogenase.Clin Chim Acta. 2019 Sep;496:18-24. doi: 10.1016/j.cca.2019.06.013. Epub 2019 Jun 12. Clin Chim Acta. 2019. PMID: 31201816
-
The local clinical validation of a new lithium heparin tube with a barrier: BD Vacutainer® Barricor LH Plasma tube.Biochem Med (Zagreb). 2017 Oct 15;27(3):030706. doi: 10.11613/BM.2017.030706. Epub 2017 Aug 28. Biochem Med (Zagreb). 2017. PMID: 28900369 Free PMC article.
-
Test results comparison and sample stability study: is the BD Barricor tube a suitable replacement for the BD RST tube?Biochem Med (Zagreb). 2020 Oct 15;30(3):030704. doi: 10.11613/BM.2020.030704. Biochem Med (Zagreb). 2020. PMID: 33071555 Free PMC article.
-
Biochemical, immunochemical and serology analytes validation of the lithium heparin BD Barricor blood collection tube on a highly automated Roche COBAS8000 instrument.Acta Biomed. 2020 Mar 19;91(1):47-55. doi: 10.23750/abm.v91i1.9195. Acta Biomed. 2020. PMID: 32191654 Free PMC article.
-
Evaluation of the BD Vacutainer(®) RST blood collection tube for routine chemistry analytes: clinical significance of differences and stability study.Biochem Med (Zagreb). 2014 Oct 15;24(3):368-75. doi: 10.11613/BM.2014.039. eCollection 2014. Biochem Med (Zagreb). 2014. PMID: 25351355 Free PMC article.
Cited by
-
Comparative Evaluation of Serum Separator V-Tube™, VQ-Tube™, and K2EDTA V-Tube™ with Becton Dickinson Tubes for Chemistry, Immunology, and Hematology Examinations.Diagnostics (Basel). 2025 Jul 14;15(14):1775. doi: 10.3390/diagnostics15141775. Diagnostics (Basel). 2025. PMID: 40722524 Free PMC article.
-
V-PRO Blood Collection Tubes: Validation for Clinical Chemistry and Immunoassay Tests.J Clin Lab Anal. 2025 Mar;39(6):e70007. doi: 10.1002/jcla.70007. Epub 2025 Mar 4. J Clin Lab Anal. 2025. PMID: 40038231 Free PMC article.
-
Evaluation of the Barricor Tube in 28 Routine Chemical Tests and Its Impact on Turnaround Time in an Outpatient Clinic.Ann Lab Med. 2021 May 1;41(3):277-284. doi: 10.3343/alm.2021.41.3.277. Ann Lab Med. 2021. PMID: 33303712 Free PMC article.
-
Serum Ferritin as a Diagnostic Biomarker for Kawasaki Disease.Ann Lab Med. 2021 May 1;41(3):318-322. doi: 10.3343/alm.2021.41.3.318. Ann Lab Med. 2021. PMID: 33303717 Free PMC article.
References
-
- Fournier JE, Northrup V, Clark C, Fraser J, Howlett M, Atkinson P, et al. Evaluation of BD Vacutainer Barricor blood collection tubes for routine chemistry testing on a Roche Cobas 8000 Platform. Clin Biochem. 2018;58:94–9. - PubMed
-
- Füzéry AK, Raizman JE, Goudreau BL, Moses K, Niemann K, Park J, et al. The BD Barricor blood collection tube is an acceptable and robust alternative to the PST for use with the Beckman AccuTnI+3 assay. Clin Biochem. 2017;50:851–7. - PubMed
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Miscellaneous