Correlation of patient characteristics with arm and finger measurements in Asian parturients: a preliminary study
- PMID: 32867679
- PMCID: PMC7457531
- DOI: 10.1186/s12871-020-01131-6
Correlation of patient characteristics with arm and finger measurements in Asian parturients: a preliminary study
Abstract
Background: Accurate blood pressure (BP) measurement depends on appropriate cuff size and shape in relation to the arm. Arm dimensions outside the recommended range of cuff sizes or trunco-conical arms may result in inaccurate BP measurements. Measuring BP using finger cuffs is a potential solution. Arm cuff size is based on mid-arm circumference (MAC), and trunco-conicity is quantified by conicity index. We aimed to determine the correlation of MAC, body mass index (BMI), and weight with conicity index.
Methods: A prospective cohort study was conducted in the KK Women's and Children's Hospital where third trimester parturients scheduled for cesarean delivery were recruited after obtaining informed consent. Parturients were asked to rate their experience with time taken to obtain BP readings, cuff popping off during measurement, need to move the cuff from the upper arm to lower arm or leg, and need to change to a different cuff. Our primary outcome was the correlation between MAC and conicity index, calculated using Pearson's correlation. The correlation between BMI and weight with conicity index was also determined.
Results: We enrolled 300 parturients. Moderate correlation was found between left MAC and left conicity index (r = 0.41, 95% CI 0.32 to 0.51), and right MAC and right conicity index (r = 0.39, 95% CI 0.29 to 0.48). Weight (r = 0.35 to 0.39) and BMI (r = 0.41 to 0.43) correlated with conicity index in this study. MAC of 1 parturient fell outside the recommended range for arm cuffs, but all parturients fit into available finger cuffs. Obese parturients had increased problems with arm cuffs popping off and needing a change of cuff.
Conclusions: BMI better correlated with conicity index compared to MAC or weight. Standard finger cuffs were suitable for all parturients studied and may be a suitable alternative.
Trial registration: Clinicaltrials.gov NCT04012151 . Registered 9 Jul 2019.
Keywords: Blood pressure; Body weight; Obstetrics.
Conflict of interest statement
Ban Leong Sng is an associate editor of BMC Anesthesiology. All other authors report no conflicts of interest in this work.
Figures
Similar articles
-
Upper limb dimensions in adults presenting for elective surgery - implications for blood pressure measurement.BMC Anesthesiol. 2020 Apr 4;20(1):76. doi: 10.1186/s12871-020-00994-z. BMC Anesthesiol. 2020. PMID: 32247309 Free PMC article.
-
Arm and finger measurements in the third trimester: Implications for blood pressure measurement.Pregnancy Hypertens. 2018 Oct;14:105-109. doi: 10.1016/j.preghy.2018.09.002. Epub 2018 Sep 5. Pregnancy Hypertens. 2018. PMID: 30527095
-
Office BP measurement using conical cuffs in children and adolescents with obesity.Blood Press. 2024 Dec;33(1):2411294. doi: 10.1080/08037051.2024.2411294. Epub 2024 Oct 11. Blood Press. 2024. PMID: 39391937
-
Arm cuff in the measurement of blood pressure.Am J Hypertens. 2002 Feb;15(2 Pt 2):67S-68S. doi: 10.1016/s0895-7061(01)02301-9. Am J Hypertens. 2002. PMID: 11866235 Review.
-
Optimal size of cuff bladder for indirect measurement of arterial pressure in adults.J Hypertens. 1989 Aug;7(8):607-13. doi: 10.1097/00004872-198908000-00002. J Hypertens. 1989. PMID: 2681407 Review.
References
-
- Knight M., Clarke B., James R. et al. on behalf of MBRRACE-UK. Saving Lives, Improving Mothers' Care - Lessons learned to inform maternity care from the UK and Ireland Confidential Enquiries into Maternal Deaths and Morbidity 2015-17. Oxford: National Perinatal Epidemiology Unit, University of Oxford 2019.