Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2021 Feb;85(2):831-844.
doi: 10.1002/mrm.28469. Epub 2020 Sep 6.

Development and validation of 3D MP-SSFP to enable MRI in inhomogeneous magnetic fields

Affiliations

Development and validation of 3D MP-SSFP to enable MRI in inhomogeneous magnetic fields

Naoharu Kobayashi et al. Magn Reson Med. 2021 Feb.

Abstract

Purpose: We demonstrate the feasibility of MRI with missing-pulse steady-state free precession (MP-SSFP) in a 4T magnet with artificially degraded homogeneity.

Methods: T1 , T2 , and diffusion contrast of MP-SSFP was simulated with constant and alternate radiofrequency (RF) phase using an extended phase graph. To validate MP-SSFP performance in human brain imaging, MP-SSFP was tested with two types of artificially introduced inhomogeneous magnetic fields: (1) a pure linear gradient field, and (2) a pseudo-linear gradient field introduced by mounting a head-gradient set at 36 cm from the magnet isocenter. Image distortion induced by the nonlinear inhomogeneous field was corrected using B0 mapping measured with MP-SSFP.

Results: The maximum flip angle in MP-SSFP was limited to ≤10° because of the large range of resonance frequencies in the inhomogeneous magnetic fields tested in this study. Under this flip-angle limitation, MP-SSFP with constant RF phase provided advantages of higher signal-to-noise ratio and insensitivity to B1+ field inhomogeneity as compared with an alternate RF phase. In diffusion simulation, the steady-state magnetization in constant RF phase MP-SSFP increased with an increase of static field gradient up to 8 to 21 mT/m depending on simulation parameters. Experimental results at 4T validated these findings. In human brain imaging, MP-SSFP preserved sufficient signal intensities, but images showed severe image distortion from the pseudo-linear inhomogeneous field. However, following distortion correction, good-quality brain images were achieved.

Conclusion: MP-SSFP appears to be a feasible MRI technique for brain imaging in an inhomogeneous magnetic field.

Keywords: brain imaging; inhomogeneous magnetic field; missing pulse steady-state free precession.

PubMed Disclaimer

Figures

Figure 1.
Figure 1.
A) Sequence diagram of 3D MP-SSFP in inhomogeneous magnetic fields (ΔB0). The sequence is composed of an RF pulse train with a constant interval, τ. Refocusing echo signals are acquired by replacing an RF pulse in every NMP (=5) RF pulses with a signal acquisition; therefore, time from the refocusing echo peak to the center of the first RF pulse in the following TR is τ (= τ1 + τ2). Gradients can be applied during excitation (light blue) and/or readout (red) to compensate the inhomogeneous field. Phase encoding is performed along two dimensions for 3D imaging. B) Extended phase graph for MP-SSFP.
Figure 2.
Figure 2.
Steady-state magnetization of MP-SSFP with constant (left column) and alternate RF phase (right column). A,B) Echo signals reaching the steady-state (9° flip angle for constant (A) and alternate (B) RF phase MP-SSFP). The refocusing echo signal intensities quickly increase at the beginning of the sequence and gradually reach the steady-state. As the missing pulse interval NMP (TR=NMP·τ) increases, stronger steady-state transverse magnetization is observed for both constant and alternate RF phase MP-SSFP. The steady-state magnetization is dependent on the flip angle (C,D). E,F) Increasing the steady-state magnetization by increasing NMP is a more time-efficient way to improve SNR than increasing the number of averages. The solid line shows SNR improvement as a continuous function proportional to T, as if averaging. The colored data points show normalized SNR compared to SNR with NMP = 3 for NMP = 3–9. The SNR advantage is more prominent with constant (E) RF phase than with alternate (F) RF phase. Simulations were conducted assuming a static gradient field of 7.6 mT/m, T1/T2 = 80/1200 ms and ADC = 1.0×10−3 mm2/s. Signal intensities were normalized with a spoiled gradient echo signal calculated with the Ernst equation (48) with TR=20 ms (5τ), flip angle = 9° and TE = 0 ms (T2* decay was ignored).
Figure 3.
Figure 3.
T1 and T2 contrast in MP-SSFP. A,B) Image contrasts are mostly T1 dependent with constant RF phase MP-SSFP; with FA = 9°, short T2 spins (<80 ms) showed stronger T2 dependence. C,D) With alternate RF phase MP-SSFP, image contrasts are relatively poor with FA = 9°, whereas signal intensities with FA = 30° are roughly dependent on T1/T2. Simulations were performed with NMP = 5, static gradient field = 7.6 mT/m and ADC = 1.0×10−3 mm2/s.
Figure 4.
Figure 4.
Diffusion contrasts due to inhomogeneous fields for MP-SSFP with constant (A-C) and alternate RF phase (D-F). To see dependence on pulse interval τ, simulation was performed with τ = 4 (A,B) and 3 ms (D,E). Interestingly, the highest steady-state signal intensity increases with constant RF phase MP-SSFP as the field gradient increases up to around 16 and 21 mT/m for τ = 4 and 3 ms, respectively (A,B). Signal intensity ratio of τ = 3 and 4 ms (Sτ=3 ms /Sτ=4 ms) demonstrates that reduced diffusion weighting with τ = 3 ms provides an SNR advantage with low flip angles as compared to τ = 4 ms (Sτ=3 ms /Sτ=4 ms > 3/4), particularly with strong background fields (C). With alternate RF phase MP-SSFP, the steady-state magnetization is highest with no background field independent of τ and decreases gradually as the field gradient increased (D,E). The SNR advantage with τ = 3 ms is more conspicuous in MP-SSFP with alternate RF phase than with constant RF phase (F). Simulations were performed with NMP = 5, T1/T2 = 1200/80 ms and ADC = 1.0×10−3 mm2/s.
Figure 5.
Figure 5.
Experimental validation of diffusion weighting. Simulations were conducted with matched T1, T2 and ADC values with an agar gel phantom (T1/T2 = 2500/80 ms and ADC = 2.0×10−3 mm2/s) for MP-SSFP with constant (A) and alternate (B) RF phase settings. Comparison of steady-state magnetization from simulation and experiments at 4 T for static field gradients of 3.7, 7.6 and 15.3 mT/m (C,D). Similar to the simulation result in Fig.4A, simulation with constant RF phase MP-SSFP showed an increase of the steady-state magnetization up to a static field gradient of 7.8 mT/m. Consistent results were observed in experiments. Similarly, the simulation results matched the experimental measurements well with alternate RF phase MP-SSFP. For lower flip angles up to 7°, the constant RF phase setting provided higher steady-state signal intensities as compared to the alternate phase setting.
Figure 6.
Figure 6.
B1+ field dependence of the MP-SSFP steady-state magnetization. A) A B1+ field of a TEM head coil was measured with the actual flip angle imaging (AFI) method at 4 T (left: sagittal and right: axial). FDA guidelines limited flip angles to under 10° due to the inhomogeneous magnetic fields used in this study. At low flip angle (nominally 6°), B1+ field inhomogeneity induces a larger variation in the steady-state magnetization using alternate RF phase MP-SSFP than it does with constant RF phase MP-SSFP (B,C). Regions with weak B1+ fields showed stronger steady-state signals with constant rather than alternate RF phase MP-SSFP (D).
Figure 7.
Figure 7.
A-F) In vivo human brain imaging with MP-SSFP in a static gradient field of 15.1 mT/m along the anterior-posterior (y) axis. Sequence parameters were: τ=3.2 ms, NMP = 5 (TR=16 ms), FA = 6.6°, HS2 pulse with a 91 kHz bandwidth and TA=3 min 43 s.
Figure 8.
Figure 8.
Image distortion correction using a B0 map obtained with MP-SSFP for a pseudo-linear field generated by mounting a head gradient 36 cm away from the 4 T magnet center. Spatial encoding was performed with phase encoding gradients along three spatial dimensions such that the reconstructed image is free from distortion from the pseudo-linear magnet field (A). A measured B0 map (B) was fitted with 4th degree polynomials (C). 3D MP-SSFP using the pseudo-linear field for frequency encoding showed conspicuous intensity and geometric distortion (D). Such image distortion was corrected using the measured B0 map (C); the distorted edge of a Teflon object in a phantom (arrows) is improved to yield a straight line in the corrected image (E).
Figure 9.
Figure 9.
In vivo human brain imaging with MP-SSFP in the pseudo-linear magnetic field. The pseudo-linear magnetic field induced severe image distortion (A). The image distortion was corrected using the B0 map shown in Fig.9C (B). The image shows anatomical structures with T1, T2 and diffusion contrasts (C-E). A region around the top of the brain shows a signal drop due to strong diffusion weighting associated with the steep field gradient in this area. Sequence parameters were: τ = 2.6 ms, NMP = 6 (TR = 15.6 ms), FA = 4°, HS2 excitation with a 140 kHz bandwidth and TA = 5 min 30 s.

References

    1. Wald LL, McDaniel PC, Witzel T, Stockmann JP, Cooley CZ. Low-cost and portable MRI. Journal of magnetic resonance imaging : JMRI 2019. - PMC - PubMed
    1. Marques JP, Simonis FFJ, Webb AG. Low-field MRI: An MR physics perspective. Journal of magnetic resonance imaging : JMRI 2019;49(6):1528–1542. - PMC - PubMed
    1. Cooley CZ, Stockmann JP, Armstrong BD, Sarracanie M, Lev MH, Rosen MS, Wald LL. Two-dimensional imaging in a lightweight portable MRI scanner without gradient coils. Magnetic resonance in medicine 2015;73(2):872–883. - PMC - PubMed
    1. McDaniel PC, Cooley CZ, Stockmann JP, Wald LL. The MR Cap: A single-sided MRI system designed for potential point-of-care limited field-of-view brain imaging. Magnetic resonance in medicine 2019;82(5):1946–1960. - PMC - PubMed
    1. Vogel MW, Guridi RP, Su J, Vegh V, Reutens DC. 3D-Spatial encoding with permanent magnets for ultra-low field magnetic resonance imaging. Scientific reports 2019;9(1):1522. - PMC - PubMed

Publication types