Evaluation of the accuracy of different apex locators in determining the working length during root canal retreatment
- PMID: 32908654
- PMCID: PMC7464225
- DOI: 10.34172/joddd.2020.026
Evaluation of the accuracy of different apex locators in determining the working length during root canal retreatment
Abstract
Background. This study aimed to assess the accuracy of three electronic apex locators (EALs) (Dentaport ZX [J Morita, Tokyo, Japan], Propex Pixi [Dentsply Maillefer, Ballaigues, Switzerland], and iPex II [NSK, Tokyo, Japan]) during root canal retreatment. Methods. The root canal lengths of 90 extracted single-rooted human teeth were determined under a dental operating microscope at ×10 magnification. The actual working length (AWL) was 0.5 mm less than the root length. Electronic measurements were performed with the three EALs. The root canals were instrumented and filled to the actual working length using the lateral compaction technique. After seven days, the teeth were retreated until the retreatment file was applied to the root canal at the working length determined by EALs, and then, the three EALs were used for determining the retreatment working length. Data were analyzed using chi-squared and Kruskal-Wallis tests. Results. In the retreatment, the accuracy of EALs was reported at %83.3 for Dentaport ZX, %83.4 forPropex Pixi, and %80 for iPex II within a tolerance of 0.5± mm of the AWL. Conclusion. Under the limitations of this study, Dentaport ZX, Propex Pixi, and iPex II can be a useful adjunct during retreatment. Clinicians should be aware that residual materials in the root canal during retreatment can affect the accuracy of EALs.
Keywords: Electronic apex locator; Retreatment; Working length.
© 2020 The Authors.
Similar articles
-
In Vivo Evaluation of 3 Electronic Apex Locators: Root ZX Mini, Apex ID, and Propex Pixi.J Endod. 2020 Feb;46(2):158-161. doi: 10.1016/j.joen.2019.10.035. Epub 2019 Dec 12. J Endod. 2020. PMID: 31839411
-
Clinical Accuracy and Precision of 3 Multifrequency Electronic Apex Locators Assessed through Micro-Computed Tomographic Imaging.J Endod. 2023 May;49(5):487-495. doi: 10.1016/j.joen.2023.02.011. Epub 2023 Feb 26. J Endod. 2023. PMID: 36848949
-
Comparative evaluation of the accuracy of electronic apex locators and cone-beam computed tomography in detection of root canal perforation and working length during endodontic retreatment.BMC Oral Health. 2024 Aug 16;24(1):953. doi: 10.1186/s12903-024-04713-9. BMC Oral Health. 2024. PMID: 39152371 Free PMC article.
-
Ex vivo accuracy of three electronic root canal length measurement devices: X-smart pro+, root ZX mini, and propex pixi.BMC Oral Health. 2025 Jul 2;25(1):1072. doi: 10.1186/s12903-025-06482-5. BMC Oral Health. 2025. PMID: 40604896 Free PMC article.
-
Influence of Coronal Preflaring on the Accuracy of Electronic Working Length Determination: Systematic Review and Meta-Analysis.J Clin Med. 2021 Jun 23;10(13):2760. doi: 10.3390/jcm10132760. J Clin Med. 2021. PMID: 34201705 Free PMC article. Review.
Cited by
-
Integrated electronic apex locators for determining working length: a systematic review of in-vitro studies.Odontology. 2025 Jul;113(3):918-929. doi: 10.1007/s10266-024-01039-4. Epub 2024 Dec 11. Odontology. 2025. PMID: 39661212
-
Effect of Different Electroconductive Root Canal Irrigations on the Accuracy of Different Apex Locators: An In Vitro Comparative Study.J Int Soc Prev Community Dent. 2024 Jun 27;14(3):211-217. doi: 10.4103/jispcd.jispcd_192_23. eCollection 2024 May-Jun. J Int Soc Prev Community Dent. 2024. PMID: 39055293 Free PMC article.
-
Evaluation of changes in root canal length and accuracy of the electronic apex locator during different stages of endodontic treatment and retreatment.Odontology. 2024 Apr;112(2):537-545. doi: 10.1007/s10266-023-00846-5. Epub 2023 Aug 29. Odontology. 2024. PMID: 37644294
-
Reliability and accuracy of dental MRI for measuring root canal length of incisors and canines: a clinical pilot study.Sci Rep. 2022 Aug 18;12(1):14068. doi: 10.1038/s41598-022-17889-3. Sci Rep. 2022. PMID: 35982139 Free PMC article.
-
Accuracy of different generations of apex locators in determining working length; a systematic review and meta-analysis.Saudi Dent J. 2022 Jan;34(1):11-20. doi: 10.1016/j.sdentj.2021.09.020. Epub 2021 Sep 25. Saudi Dent J. 2022. PMID: 35068894 Free PMC article. Review.
References
LinkOut - more resources
Full Text Sources
Miscellaneous